7050, 7050i & 7050LP
Users Manual
B-2
Summary of Error Messages, cont.
Value
Description and Corrective Action
-315
Configuration Data Lost. The configuration data has been lost. Check all
parameters to be sure they are correct.
-330
Self-Test Failed. Check the display for the test that failed.
-350
Queue Overflow. The error queue was full and messages were lost.
-400
Query Error. A read request was received when there was nothing to read.
500
Controller Malfunction. Internal control failure.
501
High Limit Exceeded. The pressure was greater than the high limit.
502
Low Limit Exceeded. The pressure was less that the low limit.
503
Slew limit Exceeded. The pressure changed faster than the slew limit
allowed.
521
Pressure Overrange. The pressure reading is outside the range of the PPI.
531
Oven Temp Overrange. Either the transistor that drives the heater for the
quartz Bourdon tube sensor (Chapter 2) or the oven temperature sensor
itself is malfunctioning. To observe the oven temperature, select OK then
select MENU/DISP. Check the transistor and sensor for malfunction,
requesting service (Chapter 7) if necessary.
533
Case Pressure Overrange. Select OK, then reduce the pressure at the case
reference port to 30 psia or lower.
540
Mechanical Zeroing Needed. The zero point of the quartz Bourdon sensor is
beyond the range of the compensation circuit. The zero is adjusted by the
software but should be manually adjusted for complete accuracy.
542
Oven Control Failure. The temperature controller is unable to keep the
sensor at the proper temperature.
545
Sensor Communication Error. Unable to Communicate with 7215xi Sensor
546
Sensor Calibration Lost. The 7215xi sensor has lost its calibration and must
be recalibrated.
600
Factory Data Lost. Internal factory constants have been lost. Contact Fluke
for more information.
601
Calibration Mode. The Calibrate button must be pressed before SCPI
calibration commands can be executed.
800
Solenoid Over-Temperature. The control solenoids have over-heated. Wait
until they have cooled before entering Control mode.
801
Error reading Barometric RPT Pressure Sensor
802
Control Sensor out of range
803
Controller Communication Error
804 Autotune
Failed
Содержание RUSKA 7050
Страница 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Страница 4: ......
Страница 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Страница 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Страница 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Страница 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Страница 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...
Страница 80: ...RUSKA 7050 7050i 7050LP Users Manual 7 4...