KnH Zero corrected counts for upper limit of range. When the zero corrected counts
from the A/D exceed this value the next higher range will be used (if possible).
(7,381,975 = full scale of current range)
KnL
Zero corrected counts for lower limit of range. When the zero corrected counts
from the A/D are below this value the next lower range will be used (if possible).
(7,381,975 = full scale of current range)
n=Range number (1 - 4 for high range sensor, 5 - 8 for low range sensor)