RUSKA 7050, 7050 & 7050LP
Users Manual
6-12
RPT — Simulated Absolute Instruments
Simulated absolute PPI’s contain two sensors that require periodic zeroing. The primary
sensor is zeroed according to the instructions in Chapter 6, Gauge and Vacuum (Negative
Gauge) Instruments, or Chapter 6, Absolute Instruments. The barometric sensor is zeroed
according to the following instructions.
1.
Verify that the Reference Port is open to atmosphere.
2.
Enter the Calibration screen by selecting MENU | CALIBRATE.
3.
Select the Atmospheric Sensor by pressing the Sensor [F6] key until the words
Atmospheric Sensor are displayed above the calibration coefficients.
4.
To begin the zeroing process, press ZERO. If the calibration access code is enabled,
enter it at the prompt. The zeroing screen will appear.
Note
The next step requires the use of a properly calibrated standard. This
zeroing process does contribute directly to the overall accuracy of the
system, since the pressure applied to the sensor is a non-zero differential
pressure.
5.
Enter the current barometric pressure. Press ENTER.
Sensor Photocell Zeroing
If the error message “Mechanical Zeroing Needed” is displayed, the sensor photocell
must be zeroed. The following steps describe this process.
WX
Warning
The PPI should only be opened by qualified
electrical/mechanical service technicians. Lethal voltages are
present and exposed in the power supply and display.
Note
The sensor MUST have zero differential between the Test port and
Reference port. For Gauge PPIs, apply atmospheric pressure to the
Reference port. For Absolute PPIs, apply a vacuum to the Reference port
(<200
μ
Hg.)
1.
Remove the PPI’s top cover.
2.
Locate and remove the plastic zeroing plug located on the side of the sensor oven
assembly.
3.
Enter the Zeroing screen by selecting MENU | CALIBRATE.
4.
Select the Primary sensor by pressing the Sensor [F6] key until the word Primary is
displayed above the calibration coefficients.
5.
Select Zero [F1]. Do NOT press the Calibrate button.
6.
Select the Mechanical Zero [F1] key.
Содержание RUSKA 7050
Страница 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Страница 4: ......
Страница 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Страница 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Страница 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Страница 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Страница 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...
Страница 80: ...RUSKA 7050 7050i 7050LP Users Manual 7 4...