Operation
Auto
Sequences
3
3-25
Select A
u
to Se
qu
ence: 1
fas42.eps
3.
Press a numeric key (
1-9
), followed by the
ENT
key.
The auto sequence begins to execute, based on the options that were enabled
using the
EDIT
function.
Note
Tests and simulations not enabled in the executing auto sequence are
skipped.
Enabled auto sequences are executed in the following order:
•
Pressure Gauge Test
•
Pressure Leak Test
•
Pressure Relief Valve Test
•
Pressure Source Test
•
Data Sheet Printout
•
NIBP Simulations
Pressure Gauge
This test monitors system pressure, which must be generated external to the Tester.
Press the
NEXT>
key to advance to the next test. Only the pressure displayed
when the
NEXT>
key is pressed appears on the printout.
Leak Test
This test performs a system leak test at the Setpoint pressure, which was defined
when the auto sequence was edited.
1.
Press the
START
soft key to initiate this test. The system then pressurizes.
Once a stable pressure reading near the Setpoint is reached, a 60-second timer
begins to count down. Once the timer has elapsed, a Leak Rate appears on the
display.
Содержание BP Pump 2 Operators
Страница 6: ......
Страница 10: ...BP Pump 2 Operators Manual iv...
Страница 12: ...BP Pump 2 Operators Manual vi...
Страница 14: ...BP Pump 2 Operators Manual 1 2...
Страница 26: ...BP Pump 2 Operators Manual 2 2...
Страница 38: ...BP Pump 2 Operators Manual 3 2...
Страница 68: ...BP Pump 2 Operators Manual...
Страница 74: ...BP Pump 2 Operators Manual B 4...
Страница 82: ...BP Pump 2 Operators Manual D 6...