![Fluke BP Pump 2 Operators Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/fluke/bp-pump-2-operators/bp-pump-2-operators_operators-manual_2310045049.webp)
Operation
Simulations
3
3-13
Note
The
<START>
soft key is not applicable to the pressure gauge test and
does not display.
3.
Apply pressure to the pressure port and read the displayed pressure in the
upper right corner of the screen.
Note
You can use a squeeze bulb or syringe to apply adequate pressure.
4.
Press
HOME
to return to the
Main
menu.
Simulations
The following sections describe various simulations that the Tester accomplishes.
For all of the following tests, it is important to select the correct cuff.
To perform simulations, take the following actions:
1.
Power on the NIBP unit.
Note
Either disable the alarm or be prepared to silence it when it sounds.
2.
Use the soft keys, beginning from the
Main
menu:
Fl
u
ke Biomedical BP B
u
mp 2
SETUP
PRESSURE
TESTS
AUTO
SEQUE
N
CE
PERFORM
SIMULATIO
N
fas31.eps
3.
Press the
PERFORM SIMULATION
soft key to access the
Select
Simulation Type
menu. From this menu, you can access the simulations
described below:
Содержание BP Pump 2 Operators
Страница 6: ......
Страница 10: ...BP Pump 2 Operators Manual iv...
Страница 12: ...BP Pump 2 Operators Manual vi...
Страница 14: ...BP Pump 2 Operators Manual 1 2...
Страница 26: ...BP Pump 2 Operators Manual 2 2...
Страница 38: ...BP Pump 2 Operators Manual 3 2...
Страница 68: ...BP Pump 2 Operators Manual...
Страница 74: ...BP Pump 2 Operators Manual B 4...
Страница 82: ...BP Pump 2 Operators Manual D 6...