![Fluke BP Pump 2 Operators Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/fluke/bp-pump-2-operators/bp-pump-2-operators_operators-manual_2310045052.webp)
BP Pump 2
Operators Manual
3-16
Patient Condition
Healthy
C
u
ff: External
OPTIO
N
S
CUFF
ready ....
120/
8
0 (93)
75 BPM
0.6
8
cc
fas08.eps
2.
Press the
OPTIONS
soft key to scroll through the simulation choices.
Parameters for Patient Condition simulations are shown in Table 3-3.
Table 3-3. Patient Condition Simulations
Patient Condition
Blood Pressure
(mmHg) (MAP)
Heart Rate
(bpm)
Pulse volume
(cc)
Healthy Heart
120/80 (93)
75
0.68
Weak Pulse
110/80 (90)
95
0.50
Mild Exercise #1
140/90 (106)
120
1.00
Strenuous Exercise #2
140/90 (106)
162
1.40
Obese Subject
120/80 (93)
90
0.50
Geriatric Subject
150/110 (123)
95
0.40
Tachycardia 120/105
(110)
130
0.40
Bradycardia 120/60
(80)
45
1.10
3.
Press the
CUFF
soft key to select Internal Adult or External cuff.
4.
Press
Start
on the NIBP unit.
5.
Compare values on the NIBP unit with those on the Tester.
6.
Press
HOME
to return to the
Main
menu.
Arrhythmias
These waveforms cause erratic readings on some NIBPMs. The blood pressure
determination strongly depends on exactly what is happening with the subject's
blood pressure when the cuff pressure is at a particular level. Some NIBPMs
pause until they detect two or more equivalent beats. The pattern of step deflations
and the measured blood pressure depend on which beats occur during each step of
the cuff pressure.
Содержание BP Pump 2 Operators
Страница 6: ......
Страница 10: ...BP Pump 2 Operators Manual iv...
Страница 12: ...BP Pump 2 Operators Manual vi...
Страница 14: ...BP Pump 2 Operators Manual 1 2...
Страница 26: ...BP Pump 2 Operators Manual 2 2...
Страница 38: ...BP Pump 2 Operators Manual 3 2...
Страница 68: ...BP Pump 2 Operators Manual...
Страница 74: ...BP Pump 2 Operators Manual B 4...
Страница 82: ...BP Pump 2 Operators Manual D 6...