BP Pump 2
Operators Manual
3-18
Respiratory Artifacts
The Respiratory Artifact exhibits a beat-to-beat variation in the blood pressure
caused by intra-thoracic pressure. Changes in the intra-thoracic pressure affect
filling of the ventricles during diastole. This in turn affects the stroke volume of
the heart. A large stroke develops a higher systolic pressure than a small stroke.
To perform Respiratory Artifact simulations, do the following:
1.
Access these simulations by pressing the
7 RESPIRATORY ARTIFACT
key. The
Respiratory Artifact
screen appears.
Respiratory Artifact
Spont Breath # 1
C
u
ff: External
OPTIO
N
S
CUFF
ready ....
13
8
/
65 (
8
9)
104 BPM
fas11.eps
2.
Press the
OPTIONS
soft key to scroll through the simulation choices.
Parameters for Respiratory Artifact simulations are shown in Table 3-5.
Table 3-5. Respiratory Artifact Simulations
Artifact Type
Blood
Pressure
(mmHg) (MAP)
Heart Rate
(bpm)
Pulse
volume (cc)
Spontaneous Breathing #1
138/65 (89)
104
Varies
Spontaneous Breathing #2
149/65 (93)
105
Varies
Spontaneous Breathing #3
112/47 (68)
86
Varies
Controlled Ventilation
132/44 (73)
98
Varies
3.
Press the
CUFF
soft key to select Internal Adult or External cuff.
4.
Press
Start
on the NIBP unit.
5.
Compare values on the NIBP unit with those on the Tester.
6.
Press
HOME
to return to the
Main
menu.
Содержание BP Pump 2 Operators
Страница 6: ......
Страница 10: ...BP Pump 2 Operators Manual iv...
Страница 12: ...BP Pump 2 Operators Manual vi...
Страница 14: ...BP Pump 2 Operators Manual 1 2...
Страница 26: ...BP Pump 2 Operators Manual 2 2...
Страница 38: ...BP Pump 2 Operators Manual 3 2...
Страница 68: ...BP Pump 2 Operators Manual...
Страница 74: ...BP Pump 2 Operators Manual B 4...
Страница 82: ...BP Pump 2 Operators Manual D 6...