Operation
Auto
Sequences
3
3-21
User-Defined
User-Defined simulations and how to create them are described in “Setting Up the
Tester.”
The following describes how to run such a simulation.
1.
Select from any of the previously-defined simulations by pressing the
0
USER DEFINED
key. The
User Defined
screen appears.
User Defined
Sim
u
lation 1:
C
u
ff: External
OPTIO
N
S
CUFF
EDIT
ready ....
120/
8
0
75 BPM
0.70 cc
fas15.eps
2.
Press the
OPTIONS
soft key to scroll through the simulation choices.
3.
Press the
CUFF
soft key to select
External
,
Internal Adult
, or
Internal
Neonate
cuff.
Note
It is important to select the correct cuff.
4.
Press
Start
on the NIBP unit.
5.
Compare values on the NIBP unit with those on the Tester.
6.
Press
HOME
to return to the
Main
menu.
Auto Sequences
It is possible to create up to nine customized auto sequences. An auto sequence
contains all four pressure tests and five simulations. The operator can disable any
of these tests or simulations. A printout of the auto sequence result can also be
enabled. The operator must make a cuff selection; this determines what NIBP
simulations are displayed for that selection.
Содержание BP Pump 2 Operators
Страница 6: ......
Страница 10: ...BP Pump 2 Operators Manual iv...
Страница 12: ...BP Pump 2 Operators Manual vi...
Страница 14: ...BP Pump 2 Operators Manual 1 2...
Страница 26: ...BP Pump 2 Operators Manual 2 2...
Страница 38: ...BP Pump 2 Operators Manual 3 2...
Страница 68: ...BP Pump 2 Operators Manual...
Страница 74: ...BP Pump 2 Operators Manual B 4...
Страница 82: ...BP Pump 2 Operators Manual D 6...