Calibrator
Menus
23
Remote Port Setup
Use the
Remote Port Setup
menu (inside the System Settings menu) to enable or disable
USBTMC, GPIB, Ethernet, and RS-232 ports by toggling a green/white switch. Tapping the
individual port buttons provides additional options and information. See the
5560A/5550A/5540A
Remote Programmers Manual
at
The menu is:
•
USB
This has USBTMC information:
•
USB0::0x0F7E::0x800A::[serial number]::INSTR
Where:
•
0x0F7E: The Fluke vendor ID
•
0x800A: The Product ID number
•
[serial number]: The serial number of the Product (the serial number is also on the rear
panel) or available using the *IDN? on one of the other remote interfaces. You can also
find it with the
Setup > About
menu.
•
Restore USB Defaults with the
Reset
button.
•
Ethernet
•
DHCP (On or Off)
•
Static IP Settings (IP Address, Gateway, Subnet Mask)
•
MAC Address
•
Port
•
End of Line Character (CR/LF, CR, LF)
•
Remote Interface (Terminal, Computer)
•
Network Security Settings (Starting Address, Ending Address)
•
Restore Ethernet Defaults
•
GPIB
•
GPIB Address
•
Restore GPIB Defaults
•
RS-232
•
Data Bits (8, 7)
•
Stop Bits (1, 2)
•
Flow Control (None, RTS/CTS, XON/XOFF)
•
Parity (None, Even, Odd)
•
Baud Rate (9600, 19200, 38400, 57600, 115200)
•
End of Line Character (CR/LF, CR, LF)
•
Remote Interface (Terminal, Computer)
•
Restore RS-232 Defaults