Calibrator
Sample Applications
59
Table 16. Adjustment- 77 Series IV DMM
Function
(Switch Position)
Adjustment Step
[1]
Input Value
E
(DC Millivolts)
C-01
600.0 mV dc
C-02
120.0 mV dc
L
(DC Volts)
C-03
6.000 V dc
C-04
60.00 V dc
C-05
600.0 V dc
C
(AC Volts)
C-06
600.0 mV, 60 Hz
C-07
600.0 V, 60 Hz
e
(Ohms)
C-08
600.0
Ω
C-09
6.000 k
Ω
C-10
60.00 k
Ω
C-11
600.0 k
Ω
C-12
6.000 M
Ω
O
(Diode Test)
C-13
5.000 V dc
mA
(Milliamps)
C-14
400.0 mA dc
C-15
400.0 mA ac, 60 Hz
A
(Amps)
C-16
6.000 A dc
C-17
6.000 A ac, 60 Hz
[1] If the DMM is not connected correctly, or if the rotary switch is in the wrong position, the DMM beeps 2 times to alert the
user.