Summary of Specifications
Specifications
A
A-3
Cycle Control Mode:
Setup time to reach initial pressure <30 seconds
Operating time to full pressure: <10 seconds (4 seconds typical)
Operating time to Min pressure: <10 seconds (7 seconds typical)
Time to complete one full cycle: <20 seconds (11 seconds typical)
Overpressure Protection: Relief valves and rupture disk set at 110% FS
Pressure port:
Test port: Autoclave F250C
Supply Port: 1/4 inch Female NPT
Display:
Graphical vacuum fluorescent
Pressure Medium:
Any non-corrosive fluid
Air Supply:
90–110 psi Dry Air (Dew Point , -10
o
C)
Peak Flow Rate of 300 SCFH
Recommended Recalibration
Interval:
1
Year
Communications:
Standard: RS-232C and IEEE-488
Syntax: SCPI (standard commands for programmable instruments)
Dimensions:
19"W x 30"D x 14"H
(49 x 76 x 36 cm)
Weight:
150
lbs
Electrical Power:
115 VAC or 240 VAC, 50/60 HZ, single phase.
Humidity:
5 to 95% non-condensing
Temperature:
Operating:
5– 50 °C (40 to 122 °F)
Storage:
-20 to 70 °C (-4 to 158 °F)
Ingress Protection:
IP20, Indoor use, Altitude <2000m
EMC:
EN 61326
Electrical Safety:
EN 61010
Pressure Safety:
Pressure equipment directive 98/23/EC class
Sound Engineering Practice (SEP)
Содержание RUSKA 7615
Страница 8: ...RUSKA 7615 Users Manual vi...
Страница 10: ...RUSKA 7615 Users Manual viii...
Страница 16: ...RUSKA 7615 Users Manual 1 6...
Страница 22: ...RUSKA 7615 Users Manual 2 6...
Страница 40: ...RUSKA 7615 Users Manual 4 12...
Страница 48: ...RUSKA 7615 Users Manual 5 8...
Страница 62: ...RUSKA 7615 Users Manual A 4...