![Fluke Calibration RUSKA 7615 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/fluke-calibration/ruska-7615/ruska-7615_user-manual_2310769047.webp)
Remote Operation
Interface Panel Emulation
5
5-7
Bit 14 Program Running.
Bit 15 0.
Questionable Status (QUES:EVENT, QUES:CONDITION, QUES:ENABLE)
Bit 0 Voltage is questionable. Set when supply voltages are not within 5%.
Bit 1 Current is questionable. Always 0.
Bit 2 Time is questionable. Set when the clock has not been set.
Bit 3. Temperature is questionable. Set when the oven temperature is not
within
range.
Bit 5. Phase is questionable. Always 0.
Bit 6 Modulation is questionable. Always 0.
Bit 7 Calibration is questionable. Set when the unit has not been calibrated.
Bit 8 Pressure is questionable. Set when the pressure is overranged.
Bit 9 Always 0.
Bit 10 Always 0.
Bit 11 Always 0.
Bit 12 Always 0.
Bit 13 Instrument Summary bit. Always 0.
Bit 14 Command Warning. Set whenever a command ignores a parameter.
Bit 15 0
Interface Panel Emulation
The 7000 may be configured to emulate the IEEE 488 command set of the RUSKA
Single Channel Interface Panel (Models 6005-701 and 6005-761). See the Interface Panel
manual for a description of the protocol. The 7000 emulation has the following
differences:
1.
The DPG is always in Remote mode (Byte 0, Bit 2)
2.
All TI strip outputs read as OFF (zero) and must be written as OFF.
3.
No special functions are implemented.
4.
Any message written to the 7000 which starts with a colon as the first character is
interpreted as a SCPI command.
To change from SCPI to Interface Panel Emulation via the remote interface send the
following message:
:SYSTem:LANGuage “6000”
To change from Interface Panel Emulation to SCPI via the remote interface send the
following message:
:SYSTem:LANGuage “SCPI”
Serial Operation
The RS-232 ports accept the same SCPI commands as the IEEE-488 port. The commands
can be terminated by a carriage return (hexadecimal 0D) or a line feed (hexadecimal 0A).
The responses are always terminated by a carriage return followed by a line feed.
The serial ports also support XON/XOFF. When the XOFF (hexadecimal 13) command
is received the Calibrator will stop transmitting. Transmission is restarted when the XON
(hexadecimal 11) command is received.
When only one unit is attached, the Control-C (hexadecimal 03) command will clear the
transmit and receive buffers and disable addressing. When addressing is disabled the unit
will respond to commands without being addressed.
Содержание RUSKA 7615
Страница 8: ...RUSKA 7615 Users Manual vi...
Страница 10: ...RUSKA 7615 Users Manual viii...
Страница 16: ...RUSKA 7615 Users Manual 1 6...
Страница 22: ...RUSKA 7615 Users Manual 2 6...
Страница 40: ...RUSKA 7615 Users Manual 4 12...
Страница 48: ...RUSKA 7615 Users Manual 5 8...
Страница 62: ...RUSKA 7615 Users Manual A 4...