RUSKA 7615
Users Manual
3-6
12.
After the pressure is stabilized press the Step
↑
or Step
↓
key. The scratchpad will be
updated with a new set-point. The step size defaults to 10% of full scale. (The step
size is set to other values in the MENU/SETUP/LIMITS screen.)
13.
Press ENTER to accept the new set-point. The controller will move to the new
set-point.
14.
Press MEASURE (the F1 key). The Calibrator will change to Measure mode. No
confirmation is necessary to leave Control mode.