1586A
Calibration Manual
36
Table 10. Adjustment Steps (cont.)
Step
Value Range
Input
Signal
Description
Entry
Point
DCI
DC100UA 100.0E-6
Start calibration of dc
100
μ
A range
Y
DC_P100UA_1
100.0
μ
A
N
DC_0A_1
Open
N
DC_N100UA_1
-100.0
μ
A
N
DC1MA 1.0E-3
Start calibration of dc 1 mA
range
Y
DC_P1_0MA_2
1.0
mA
N
DC_0A_2
Open
N
DC_N1_0MA_2
-1.0
mA
N
DC10MA 10.0E-3
Start calibration of dc
10 mA range
Y
DC_P10MA_3
10.0
mA
N
DC_0A_3
Open
N
DC_N10MA_3
-10.0
mA
N
DC100MA 100.0E-3
Start calibration of dc
100 mA range
Y
DC_P100MA_4
100.0
mA
N
DC_0A_4
Open
N
DC_N100MA_4
-100.0
mA
N