SUPER-DAQ Precision Temperature Scanner
Required Equipment
19
Table 4. Required Test Equipment (cont.)
Function Instrument
Type
Model
Comments
Current DC
Standard Fluke
5522A
Must be characterized with
8508A
8½ digit meter
Fluke 8508A
Used to characterized the
5522A
Alternate standard
[1]
Fluke
5720A
1586-2586
Input Module
CJC
Metrology Drywell
Fluke 9171 with insert
block
Reference
thermistor probe
Fluke 5610-9
Accuracy 0.013
°
C or
better at 25
°
C
Reference
thermometer
Fluke 1586A
Only for CJC verification
E-type
thermocouples
Omega
TT-E-24-SLE
Must be calibrated.
Accuracy 0.026
°
C or
better at 25
°
C.
1586-2588
DAQ-STAQ
CJC
Metrology Drywell
Fluke 9171 with insert
block
Reference
thermistor probe
Fluke 5610-9
Accuracy 0.013
°
C or
better at 25
°
C.
Reference
thermometer
Fluke 1586A
Only for CJC verification
E-type
thermocouples
Omega
TT-E-24-SLE with NMP-
E-M TC adaptor
Must be calibrated.
Accuracy 0.026
°
C or
better at 25
°
C.
Cables
To reduce the possibility of inducing errors with ac signals picked up by the test
leads, use short, shielded twisted-pair PTFE-insulated test cables between the test
equipment and the Meter. Fluke makes a 2 foot (PN 738716) and 4 foot (PN
738724) PTFE insulated test cable for this purpose
[1]
Other alternate standards beside those listed can be used as long as they provide sufficient traceable [Test
uncertainty Ratios (TURs)] at each calibration and verification point.