Required Equipment
Table 4 lists the equipment required for performance tests and calibration of the
Product.
Table 4. Required Test Equipment
Function Instrument
Type
Model
Comments
Volts DC
Standard Fluke
5720A
8½ digit meter
Fluke 8508A
Used to characterized the
5522A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
[1]
Fluke
5522A
Must be characterized with
8508A
Ohms
Standard Fluke
5720A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
The standard resistors
which have equivalent
spec with Fluke 5720A.
PRT
Standard
Fluke 5522A (for 1 k
Ω
value)
Standard resistors of
25
Ω
, 100
Ω
, 200
Ω
and
400
Ω
Resistance accuracy
specification (k = 2) should
be equal to or better than:
6 ppm (for 25
Ω
)
4 ppm (for 100
Ω
, 200
Ω
and 400
Ω
)
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
The standard resistors
which have equivalent
specification
Thermistor
Standard Fluke
5522A
Must be characterized with
8508A
8½ digit meter
Fluke 8508A
Used to characterized the
5522A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
[1]
Fluke
5720A