1586A
Calibration Manual
34
Table10. Adjustment Steps (cont.)
Step
Value Range
Input
Signal
Description
Entry
Point
OHM
R100 100.0
Start calibration of FRES
100
Ω
range
Y
R_100_1
100.0
Ω
N
R_19_1
19.0
Ω
N
R_0_1
0.0
Ω
N
R1K 1.0E+3
Start calibration of FRES 1
k
Ω
range
Y
R_1K_2
1.0
k
Ω
N
R_190_2
190.0
Ω
N
R_0_2
0.0
Ω
N
R10K 10.0E+3
Start calibration of FRES 10
k
Ω
range
Y
R_10K_3
10.0
k
Ω
N
R_1_9K_3
1.9
k
Ω
N
R_0_3
0.0
Ω
N
R100K 100.0E+3
Start calibration of FRES
100 k
Ω
range
Y
R_100K_4
100.0
k
Ω
N
R_19K_4
19.0
k
Ω
N
R_0_4
0.0
Ω
N
R1M 1.0E+6
Start calibration of FRES 1
M
Ω
range
Y
R_1M_5
1.0
M
Ω
N
R_190K_5
190.0
M
Ω
N
R_100K_5
100.0
k
Ω
N
R_0_5
0.0
Ω
N
R10M 10.0E+6
Start calibration of RES 10
M
Ω
range
Y
R_10M_6
10.0
M
Ω
N
R_1_9M_6
1.9
M
Ω
N
R_1_0M_6
1.0
M
Ω
N
R_0_6
0.0
Ω
N
Содержание 1586A Super-DAQ
Страница 6: ...1586A Calibration Manual iv...
Страница 8: ...1586A Calibration Manual vi...
Страница 54: ...1586A Calibration Manual 46 1 2 3 4 6 5 7 8 hcn209 eps Figure 10 Replaceable Parts...