SUPER-DAQ Precision Temperature Scanner
Performance Tests
27
Table 9. DC Current Verification Steps
Nominal
Input (mA)
Range
1-Year Test Limits
High Low
0
0.1
3.5 nA
-3.5 nA
0.1
[1]
0.1
100.0185
μ
A 99.9815
μ
A
-0.1
[1]
0.1
-99.9815
μ
A -100.0185
μ
A
0
1
11.0 nA
-11.0 nA
1
[1]
1
1.000161 mA
0.999839 mA
-1
[1]
1
-0.999839 mA
-1.000161 mA
0 10
0.35
μ
A -0.35
μ
A
10
[1]
10
10.00185 mA
9.99815 mA
-10
[1]
10
-9.99815 mA
-10.00185 mA
0 100
3.5
μ
A -3.5
μ
A
100
[1]
100
100.0185 mA
99.9815 mA
-100
[1]
100
-99.9815 mA
-100.0185 mA
[1]
5720A or 5522A must be used with 8508A to obtain suitable test uncertainty ratio.
1586-2586 CJC Module Accuracy Verification
1. Connect the calibrated E-type thermocouples to channel 10 on the Product
(UUT).
2. Insert the thermocouples into a drywell calibrator which is set and stabilized
at 25
°
C.
3. Configure the channels of the Product for an E-thermocouple.
4. Insert the reference thermistor probe into the drywell to measure the actual
temperature.
5. Use the reference thermometer with the reference thermistor probe and the
thermocouples with the Product to measure the drywell temperature until all
readings are stable.
6. Compare the thermocouple readings and reference thermometer reading.
7. The difference should be <0.6
°
C.
Содержание 1586A Super-DAQ
Страница 6: ...1586A Calibration Manual iv...
Страница 8: ...1586A Calibration Manual vi...
Страница 54: ...1586A Calibration Manual 46 1 2 3 4 6 5 7 8 hcn209 eps Figure 10 Replaceable Parts...