Calibration Manual
24
PRT Verification
Connect the Product as in Figure 5 for the 1 k
Ω
value listed in Table 7. For all
other values in the table, use standard resistors. Configure Ch001 as a 4-wire
PRT function with resistance display.
5522A Calibrator
1 k
e
UUT
NC
NC
hcn312.eps
Figure 5. PRT Test Equipment Setup
Table 7. PRT Verification Steps
Nominal
Input
Range
1-Year Test Limits
High Low
0
Ω
100
0.0008
Ω
-0.0008
Ω
25
Ω
100
25.0008
Ω
24.9992
Ω
100
Ω
100
100.002
Ω
99.998
Ω
1 k
Ω
1k
1.00006
k
Ω
0.99994
k
Ω