32
Use [
ESC
] to go back to the previous menu
– changes will not be adopted.
The default node ID is 1.
5.3.3.
Start
For starting the controlling with CAN use the [
] buttons to select the item
START
.
Confirm by hitting the [
] button.
Now all parameter settings can be made with the CAN-protocol which is described in
the following.
When the device is switched-off during the CAN-mode, it goes in this mode again after
a reset.
Use [
ESC
] to leave the CAN-mode.
After that there is the
initialisation
phase
,
followed
from the
Bootup
–
message
. This
message has the following buildup:
Identifier
Data
700h + Node ID
00h
If this message
could not be sent
there
is a message “
TRANSMIT
ERROR
” on
the display for 1 second.
After
this message the state
Pre-
Operational
is achieved.
In the menu bar you can see a “
c
”.
This is the sign that the device is now
controlled with CAN. The state of the
device is shown on the display.
As soon as the device is put into the state Operational, measurement is started. In this
state the contamination classes are shown on the display according to the chosen
report format.
5.4. Object dictionary
The object dictionary has a buildup like a table with lines and columns.
An entry of the object dictionary is addressed with an index and a subindex.