2.8.3.3.1. Limits
When achieving or exceeding the set limit values in the measuring cycle the corresponding
threshold contacts switch or rather close.
With the setting of zero values the functionality of the threshold contacts is switched off
no switching or rather closing of the threshold contacts.
Threshold settings can be changed for each classification type and each counting
channel.
These settings will also be stored in the back-up RAM.
Use the [
] buttons to select the type of classification.
The selected type will be marked with a
*
on the left side of the menu.
Go to the selected classification type by hitting [
].
Classification ISO 4406:99:
Use the [
] buttons to choose the counting channel that has to be changed and
confirm by hitting [
]. Use [
] to set new contamination class threshold digit by digit
and confirm with [
].
If new contamination class thresholds are set for all necessary counting channels use
[
ESC
] to go back to the previous menu.
Classification NAS 1638:
Use the [
] buttons to choose the counting channel that has to be changed and
confirm by hitting [
].
Use [
] to set new contamination
class threshold digit by digit and
confirm with [
].
If new contamination class thresholds are set for all necessary counting channels use
[
ESC
] to go back to the previous menu.