13
2.8.3.1.3. Start
Measurements are performed automatically and the results can be displayed as
contamination classes. (depending upon selected classification type)
During the measuring the particles per counting channel and the volume flow constantly
put out via the RS232 interface. (see chapter 3.2)
Serially, the CCM 01 comes with the data manager software. This software edits and
manages measurements stored by the CCM 01.
For this application a storage interval has to be selected. (see chapter 2.8.3.1.4)
Measurements will be stored continuously on the internal memory of the CCM 01
according to the selected storage interval. They can be transferred to an external
computer using the RS232 interface or the Bluetooth modul EBT 01 in order to be
edited and managed using the data manager software. (see chapter 3.3 )
Fixed limiting values are shown by a rectangle.
Exceeded limiting values are shown by a filled out rectangle in the display.
After disconnecting the power supply during the measurement it will
automatically continue the measurement after reconnecting.
Measurements will be performed until the operator hits [
ESC
] and switches back to the
[
MEASURE
] menu.
2.8.3.1.4. Storage Interval
Use the [
] button to choose storage interval.
The
intervals
that can be
chosen
are between
five minutes
and
seven days
.
Storing
the set interval will occur in a
battery backed RAM
.
In case of a power outage
,
the
setting is saved
.
100 measurements
can be stored. After that, the oldest measurement is overwritten.
Chose the storage interval with the [
] buttons and confirm by hitting the [
] button
–
automatic return to the [
MEASURE
]
– menu.
Use [
ESC
] to go back to the previous menu
– changes will not be adopted.