6. Rotate the inner ring of the lens 30° clockwise. The lens makes a click when it locks
in place.
7. Make sure that the two index marks are aligned, indicating that the lens is locked in
place.
7.15
Calibrating the lens–camera combination
7.15.1
Introduction
Before a new lens can be used with the camera, the lens–camera combination must be
calibrated
.
This is a process that previously had to be performed by a FLIR service department, but
for the FLIR T5xx series the calibration can be performed by the user. This feature is
called AutoCal. The AutoCal procedure requires a calibration target, which is included in
the lens package.
Note
For lenses that ship with the camera, the lens–camera combination is factory
calibrated.
#T810253; r. AL/84068/84441; en-US