UMPIOXR400V2-7-1EN, 2012-07-01
61
14 Inputs (Option)
Assign the corresponding process values to the the minimum and maximum value of the input signal.
If
NO DETECTION
is selected skip the following two displays.
Enter the maximum value of the input signal (here:
20 mA
).
This value is given in the specification of the external transducer.
Press ENTER.
Enter the process quantity (here:
DENSITY
) to be assigned to the min-
imum value of input signal (here
4.0 mA
).
This value is given in the specification of the external transducer.
Press ENTER.
Enter the process quantity (here:
DENSITY
) to be assigned to the
maximum value of input signal (here
20.0 mA
).
This value is given in the specification of the external transducer.
Press ENTER.
Define the behavior of the external transducer in case of error. The fol-
lowing items can be selected:
•
NO DETECTION
: The external transducer does not evaluate the
signal before sending it to the transmitter.
•
If I1 < MIN (4.0 mA)
: An error is sent if the value falls below
the previously defined measuring range.
•
If I1 > MAX (20.0 mA)
: An error is sent if the value exceeds
the previously defined measuring range.
Note!
The error behavior of the external transducer is described in the specification.
If I1 < MIN (4.0 mA)
is selected:
Enter the lowest value for the measuring range before an error mes-
sage is sent.
This value is given in the specification of the external transducer.
If I1 > MIN (20.0 mA)
is selected:
Enter the highest value for the measuring range before an error mes-
sage is sent.
This value is given in the specification of the external transducer.
The terminals to be used for the connection of the inputs are now dis-
played. Press ENTER.
I 1 : S e n s o r M A X
2 0 . 0 0 m A
I 1 : M I N ( 4 . 0 m A ) =
1 . 0 0 0 0 0 g / c m 3
I 1 : M A X ( 2 0 . 0 m A ) =
1 . 2 0 0 0 0 g / c m 3
E r r o r
d e t e c t i o n
N o d e t e c t i o n
m A
E r r o r i f I 1 <
3 . 0 m A
m A
E r r o r i f I 1 >
2 1 . 0 m A
m A
I n p u t I 1
T e r m i n a l : T 1 A , B / b