10 Serial Output
10.2 Format of the Serial Output
At the beginning of the measurement, a header describing the format of the output table is sent. The measured val-
ues are transmitted afterwards. In standard format, the columns are separated by a semicolon. Each line ends with
a line feed. The first three lines of the header contain general information about the instrument and the measure-
ment (see example below):
Configuration parameters output for each measuring channel in a data block are in the following lines.
The line
\DATA
will be transmitted next.
\DEVICE
serial number of the instrument and firmware version
\STORAGE DATE
date and time of the measurement
\STORAGE RATE
output rate (time between two data lines belonging to the same channel)
\A:-;CHANNEL_INFORMATION
information about the channel (designation of the measuring point, name of the
fluid, selected physical quantity). Each line begins with the name of the channel
(
\A:
,
\B:
)
\A:=;COLUMN DESCRIPTIONS
The content of each column sent for this channel is specified on one line:
- column number (
COL_01, COL_02
, ...)
- source item
- unit of measurement
- column title
- information
Example:
\DEVICE ;PXTR372-07200101(V1.92)
\STORAGE_DATE;12.08.2004 14:51:46
\STORAGE_RATE;3 sec
\A:-;CHANNEL_INFORMATION
\A:-;Measuring_point=;123-00-33
\A:-;Bank[0].Main.Fluid=;MyOwnSubstance;Scale:Brix
\A:-;COLUMN DESCRIPTION
\A:-;COL_01=;Mass-Concentr.;[% mas];(MME);
\A:-;COL_02=;StatusBit R1[bit];(R1S);Measured values/Main values
\A:-;COL_03=;Fluid temperat.;[C];(SFT);
\A:-;COL_04=;Fluid pressure[?];(LPR);UNLINKED!
\A:-;COL_05=;StatusBit R3[bit];(R3S);Sensor values/Refr.index
\A:-; (MME)[% mas]; (R1S)[bit]; (SFT)[C]; (LPR)[?]; (R3S)[bit];
\DATA
A:;
10.26; NOT; 22.13; ?E?; NOT;
A:;
10.26; NOT; 22.12; ?E?; NOT;
A:;
10.26; NOT; 22.12; ?E?; NOT;
A:;
10.27; TRIG; 22.11; ?E?; NOT;