64
UMPIOXR400V2-7-1EN, 2012-07-01
A Technical Data
measuring functions
physical quanti-
ties
refractive index, fluid temperature, °Brix, w% (saccharose, fructose, glucose, invert sugar), further by RMKoeff
physical quanti-
ties for
diagnosis
signal amplitude, sensor humidity, sensor temperature
communication
interface -
process integration:
RS485 (option)
- diagnosis: RS232
- process integration:
RS485 (option)
- diagnosis: RS232
1
- process integration:
RS485 (option)
- diagnosis: RS232
- process integration:
RS485 (option)
- diagnosis: RS232
- process integration:
RS485 (option)
- diagnosis: RS232
serial data kit
software (all Win-
dows
TM
ver-
sions)
RMKoeff: management of fluid data sets
cable RS232
RS232
1
RS232
RS232
RS232
adapter
RS232 - USB
RS232 - USB
1
RS232 - USB
RS232 - USB
RS232 - USB
outputs (option)
The outputs are galvanically isolated from the transmitter.
number on
request
current output
range 0/4...20
mA
accuracy
0.1 % of reading ± 15
A
active output
R
ext
< 500
passive output
U
ext
= 4...24 V, dependent on R
ext
R
ext
< 1 k
voltage output
range
0...1 V or 0...10 V
accuracy
0...1 V: 0.1 % of reading ± 1 mV
0...10 V: 0.1 % of reading ± 10 mV
internal re-
sistance
R
i
= 500
frequency output
range
0...1 kHz or 0...10 kHz
open collector
24 V/4 mA
binary output
open collector
24 V/4 mA
24 V/4 mA
Reed relay
48 V/0.25 A
48 V/0.25 A
optorelay
32 V/100 mA
-
binary output as
alarm output
- functions
limit
limit
inputs (option)
The inputs are galvanically isolated from the transmitter.
number
max. 4, on request
current input
range
active: 0...20 mA
passive: -20...+20 mA
accuracy
0.1 % of reading ± 10
A
active input
U
i
= 24 V, R
i
= 50
, P
i
< 0.5 W,
not short-circuit proof
passive input
R
i
= 50
, P
i
< 0.3 W
voltage input
range
0...1 V or 0...10 V
accuracy
0...1 V: 0.1 % of reading ± 1 mV
0...10 V: 0.1 % of reading ± 10mV
internal re-
sistance
R
i
= 1 M
1
connection of the interface RS232 outside of explosive atmosphere (housing cover open)
transmitter
PIOX
TR374
TR374A2
TR374B
TR374C
TR379