UMPIOXR400V2-7-1EN, 2012-07-01
11 Output Options
11.4.3 Selection of the Source Item
The source item is the quantity triggering the status bit, i.e the quantity about which information is to be sent to the
status output.
Example:
PIOX R400 will be used to control the Brix value of a sugar solution. The value must not exceed
30 °. Select the Brix value as source item for the status output. PIOX R400 will monitor the Brix
value and raise an alarm when the defined limit is not respected. The alarm can be sent to the pro-
cess control system.
Select the source item type in the scroll list. The different source item
types and source items are given in Tab. 11.1. Press ENTER.
If the source item type
MEASURED VALUES
is selected, select the
source item in the scroll list. Press ENTER.
Tab. 11.1: Available source items for the status outputs
source item type
possible source items
MEASURED VALUES
MAIN MEASURE
main physical quantity selected in program branch PARAMETER
AUXILARY VALUE
auxiliary physical quantity
NORMALIZED nD
normalized refractive index
SENSOR VALUES
REFRACTIVE INDEX
measured refractive index (not normalized)
FLUID TEMPERAT.
fluid temperature measured by the sensor
SENSOR TEMPERAT.
average inner temperature of the sensor
HUMIDITY
sensor humidity
SIGNAL AMPLITUDE
relative signal amplitude
SIGNAL QUALITY
relative signal quality
PIXEL POSITION
absolute pixel position
ZERO POSITION
absolute zero point of the pixels
LINKED INPUTS
FLUID TEMP.
fluid temperature
FLUID PRESSURE
fluid pressure
DENSITY
density
KINEM. VISCOSITY
kinematic viscosity
DYNAM.VISCOSITY
dynamic viscosity
CONCUCTANCE
conductance
pH VALUE
pH value
SOUND SPEED
sound velocity
VOLUME FLOW
volume flow
COMMONINP x
with x = 1...4, any output
If the source item type
LINKED INPUTS
is selected, select the source
item in the scroll list. Press ENTER.
R 1 : S o u r c e I t e m
M e a s u r e d v a l u e s
M e a s u r e d
v a l u e s
M a s s - C o n c e n t r .
L i n k e d i n p u t s
? F l u i d t e m p .