17 SuperUser mode
FLUXUS G809
17.6 Detection of long measurement failures
153
UMFLUXUS_G809V2-2EN, 2020-06-25
17.6
Detection of long measurement failures
If there are no valid measured values during a long time interval, new increments of the totalizers will be ignored. The
values of the totalizers remain unchanged.
In the SuperUser mode, it is possible to set the time interval.
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Do not total. if no meas.
is displayed.
• Enter the time. If zero is entered, the default value of 30 s will be used.
• Press ENTER.
17.7
Number of decimal places of the totalizers
The totalizer values can be displayed with up to 11 places, e.g.,
74890046.03
. In the SuperUser mode, it is possible to
define the number of decimal places.
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Total digits
is displayed.
• Select one of the following list items:
–
Automatic
: dynamic adaptation
–
Fixed to x digit
: x decimal places (range: 0...4)
• Press ENTER.
Total digits = Automatic
The number of decimal places will be adjusted dynamically. Low totalizer values will initially be displayed with 3 decimal
places. If the values of the totalizers are higher, the number of decimal places will be reduced.
Total digits = Fixed to x digit
The number of decimal points is constant. The max value of the totalizer is reduced with the number of decimal places.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Do not total. if no meas.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Total digits
max. value
display
< 10
6
±0.000
…
±999999.999
< 10
7
±1000000.00
…
±9999999.99
< 10
8
±10000000.0
…
±99999999.9
< 10
10
±1000000000
…
±9999999999
decimal places
max. value
max. display
0
< 10
10
±9999999999
1
< 10
8
±99999999.9
2
< 10
7
±9999999.99
3
< 10
6
±999999.999
4
< 10
5
±99999.9999
Notice!
The number of decimal places and the max. value defined here only affect the display of the totalizers.