9  Measurement

FLUXUS G809

 9.1  Parameter input

89

UMFLUXUS_G809V2-2EN, 2020-06-25

Pipe wall thickness

• Enter the pipe wall thickness. 
• Press ENTER.

Pipe material

The pipe material has to be selected to be able to determine the corresponding sound speed.
The sound speeds for the materials in the scroll list are stored in the transmitter.
• Select the pipe material.
• Press ENTER.
• If the material is not in the scroll list, select 

Other Material

.

• Press ENTER.

Sound speed of the pipe material

• Enter the sound speed of the pipe material.
• Press ENTER.

These displays will only be indicated if 

Other Material

 is selected.

For the sound speed of some materials, see annex C.

Lining

• Select 

yes

 if the pipe has a lining. Select 

no

 if the pipe has no lining.

• Press ENTER.

Lining material

• Select the lining material.
• Press ENTER.
• If the lining material is not included in the scroll list, select 

Other Material

.

• Press ENTER.
This display will only be indicated if 

yes

 is selected in the menu item 

Lining

.

Sound speed of the lining material

• Enter the sound speed of the lining material.
• Press ENTER.

Parameter\Wall Thickness

Parameter\Pipe Material

Parameter\Pipe Material\Other Material\c-Material

Notice!

There are 2 sound speeds for pipe materials, the longitudinal and the transversal one. Enter the sound speed which 

is nearer to 2500 m/s.

Parameter\Lining

Parameter\Lining

Parameter\Lining\Other Material\c-Material

Содержание FLUXUS G809

Страница 1: ...FLUXUS G809 UMFLUXUS_G809V2 2EN Operating instruction...

Страница 2: ...660 Fax 49 30 936 67 680 E mail info flexim com www flexim com FLUXUS is a registered trademark of FLEXIM GmbH Operating instruction for FLUXUS G809 UMFLUXUS_G809V2 2EN 2020 06 25 Article number 2167...

Страница 3: ...enetration 16 3 4 Undisturbed flow profile 16 3 5 Influence of noise 18 3 6 Selection of the measuring point taking into account the flow profile and the influence of noise 19 4 Product description 21...

Страница 4: ...Calibration 115 11 4 Examination of the O ring 115 11 5 Examination of the thread on the flameproof enclosure housing 116 11 6 Examination of housing and sight glass 116 12 Dismounting and disposal 11...

Страница 5: ...erUser mode 148 17 1 Activation deactivation 148 17 2 Transducer parameters 149 17 3 Defining flow parameters 149 17 4 Limit of the signal amplification 151 17 5 Upper limit of the sound speed 152 17...

Страница 6: ...Table of contents FLUXUS G809 2020 06 25 UMFLUXUS_G809V2 2EN 6...

Страница 7: ...ion or miss information please inform us We will be grateful for any suggestions and comments regarding the concept and your experience working with the mea suring equipment If you have any suggestion...

Страница 8: ...l of electrical equip ment 2 2 Intended use The measuring equipment is intended for the measurement of fluid properties in closed pipes By means of connected transducers the transit times of the ultra...

Страница 9: ...ly be replaced by FLEXIM The operator shall carry out periodic checks for changes or damages that can present a danger For further information contact FLEXIM Observe the specifications for the install...

Страница 10: ...operly packed for transport Use if possible the original packaging by FLEXIM or an equivalent cardboard box Position the transmitter transducers and accessories in the middle of the cardboard box Fill...

Страница 11: ...w in the pipe is mostly turbulent Transition range The flow is partly laminar and partly turbulent Sound speed c Speed of the propagating sound The sound speed depends on the mechanical properties of...

Страница 12: ...ocity The signals are emitted and received by 2 transducers alternatively in and against the flow direction If the fluid is flowing the signals propagating in the fluid are displaced with the flow Cau...

Страница 13: ...e signal against the flow direction c sound speed 1 transducer emitter 2 transducer receiver 3 pipe wall Fig 3 3 Transit time difference t 1 signal in the flow direction 2 signal against the flow dire...

Страница 14: ...e the transducer emitting the ultrasonic signal and the transducer receiving it One beam consists of 1 or several sound paths Transducer distance The transducer distance is measured between the inner...

Страница 15: ...3 7 2 sound paths in 1 plane diagonal arrangement with 1 beam reflection arrangement with 1 beam 1 transducer pair 1 sound path 1 beam 1 plane 1 transducer pair 2 sound paths 1 beam 1 plane diagonal a...

Страница 16: ...pipe s upper part are pre vented from influencing the propagation of the signal see Fig 3 8 and Fig 3 9 3 4 Undisturbed flow profile Some flow elements e g elbows valves pumps reducers distort the flo...

Страница 17: ...40 D outlet l 5 D disturbance source valve inlet l 40 D outlet l 5 D disturbance source reducer inlet l 10 D outlet l 5 D disturbance source compressor inlet l 20 D Tab 3 1 Recommended distance from...

Страница 18: ...the same time Measuring points to be avoided measuring point directly at the reflection point l 3 D measuring point at a distance of lS 2 D from the reflection point pipe wall signal and measuring sig...

Страница 19: ...file as possible Example fluid natural gas c 400 m s pipe material stainless steel c 3000 m s length of pipe segment 1 20 D length of pipe segment 2 20 D number of sound paths 2 ls 7 5 D area with dev...

Страница 20: ...plete pipe segment 2 area with low influence of noise reflection point flange recommended area for the measuring point l 3 D and outside of lS 7 5 2 D on pipe segment 1 Fig 3 13 Area for the measuring...

Страница 21: ...easured values can be displayed used for calculations and transmitted by the transmitter 4 2 Handling concept The command panel is on the front side of the housing The keys are operated by a magnetic...

Страница 22: ...he lower line Press key to scroll and select a list item in the upper line Press key to scroll and select a value for the selected list item in the lower line Press ENTER 4 3 2 Input fields Enter the...

Страница 23: ...ement by inadvertently pressing key BRK Tab 4 3 Navigation scroll to the right or up through a scroll list scroll to the left or down through a scroll list Tab 4 4 Input of numbers move the cursor to...

Страница 24: ...aging the transmitter can fall down There is a danger of crushing body parts or damaging the measuring equipment Secure the transmitter against falling during packaging Wear the required personal prot...

Страница 25: ...may result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Installation connection...

Страница 26: ...make sure that the housing is closed correctly and that the set screws are tightened see Fig 6 1 Caution Touching hot or cold surfaces This may result in injuries e g thermal damages Observe the ambi...

Страница 27: ...the transmitter to the instrument mounting plate Fix the transmitter with the shackle to the pipe by tightening the nuts 3 Installation on a pipe 2 The instrument mounting plate 2 is fixed to the pip...

Страница 28: ...6 Installation 6 1 Transmitter FLUXUS G809 2020 06 25 UMFLUXUS_G809V2 2EN 28 Fig 6 3 Installation on the pipe 1 shackle 2 instrument mounting plate 3 nut 4 tension strap 1 2 3 2 4...

Страница 29: ...ansducer cable range The ambient temperature at the measuring point has to be within the operating temperature range of the transmitter and the transducers see technical specification If the measuring...

Страница 30: ...nge ment transverse flow effects are compensated because the beam crosses the pipe in 2 directions higher accuracy of measurement be cause the accuracy increases with the number of sound paths Diagona...

Страница 31: ...ected in an angle of 90 45 to the elbow plane The elbow is upstream of the measuring point Bidirectional measurements Measurements in reflection arrangement with 2 beams and 2 planes The sound beam pl...

Страница 32: ...n the pipe lengthwise or crosswise They are mounted on the entire pipe circumfer ence Pipe damping mats can be mounted to reduce the propagation of acoustic noise in the pipe wall if it is not possibl...

Страница 33: ...ng mat see 1 in Fig 6 7 b Then move the roller in the middle of the damping mat along the pipe circumference see 2 in Fig 6 7 b Repeat the steps until all damping mats are fixed to the pipe The dampin...

Страница 34: ...ion point 1 reflection point 2 reflection points 2 reflection points 1 transducer mounting fixture 2 transducer damping mats 3 pipe damping mats 4 reflection point lS1 mounting length of the transduce...

Страница 35: ...otal mounting length is 1010 mm Fig 6 8 Outer pipe diameter 900 mm 1 transducer damping mats 2 pipe damping mats 3 reflection point The transducer damping mats are mounted lengthwise on the pipe The p...

Страница 36: ...reflection point 2 reflection points 2 reflection points 1 transducer mounting fixture 2 transducer damping mats 3 pipe damping mats 4 reflection point lS1 mounting length of the transducer damping ma...

Страница 37: ...nsducer damping mat lS1 600 mm pipe damping mat lR 600 mm 2 The total mounting length is 1800 mm Fig 6 9 Outer pipe diameter 900 mm 1 transducer damping mats 2 pipe damping mats 3 reflection point The...

Страница 38: ...ble during installation it has afterwards to be remounted and fixed to the transducer cable using the provided cable tie The heat shrink tubing must not be reused Fig 6 11 Transducer arrangement in mo...

Страница 39: ...UMFLUXUS_G809V2 2EN 2020 06 25 6 2 3 4 Mounting with Variofix L PermaRail Scope of delivery 2 Variofix L 1 transducer pair 4 quick release clasp with tension strap or 4 band clamp clasp with tension s...

Страница 40: ...nsducer mounting fixtures in reflection arrangement is described 1 transducer mounting fixture for each transducer Overview of the mounting steps step 1 dismounting of the transducer mounting fixture...

Страница 41: ...n instructions according to the supplied clasp Band clamp clasp The clasp is fixed to the tension strap see Fig 6 16 Quick release clasp The clasp is fixed to the tension strap see Fig 6 17 Cut the te...

Страница 42: ...nstallation instructions according to the supplied clasp Band clamp clasp Insert the tension strap into the tension strap clamp see Fig 6 20 Position the clasp and the tension strap clamp on the pipe...

Страница 43: ...r pipe diameter 80 mm pipes that are not subjected to significant temperature fluctuations Position the ratchet clasp tension strap clamp and metal spring if necessary on the pipe see Fig 6 19 on a ho...

Страница 44: ...on strap clamp 1 see Fig 6 26 Tighten the nut of tension strap clamp 1 but not too firmly in order not to damage the tension strap Notice In order to release the screw and the tension strap press the...

Страница 45: ...ion strap see Fig 6 27 Quick release clasp Insert the tension strap into the tension strap clamp 2 and the metal spring see Fig 6 29 and Fig 6 27 Place the tension strap around the pipe and insert it...

Страница 46: ...rough the slot of the clasp screw see Fig 6 31 Tighten the tension strap Cut off the protruding tension strap see Fig 6 31 Tighten the clasp screw Tighten the nut of the tension strap clamp 2 but not...

Страница 47: ...train relief clamp to protect them from mechanical strain see Fig 6 34 Stick coupling foil or some coupling compound for a short term installation on the contact surface of the transducers The couplin...

Страница 48: ...clamp 3 equipotential bonding terminal a transducer distance Notice Make sure that the coupling foil remains on the contact surface of the transducers For information concerning the coupling foil see...

Страница 49: ...ides of the pipe see Fig 6 37 In the following the installation of 1 transducer mounting fixture in reflection arrangement is described Overview of the mounting steps step 1 dismounting of the transdu...

Страница 50: ...t it off see Fig 6 39 Step 2 Mounting the rail Select the installation instructions according to the supplied clasp Mounting of the rail without a clasp Cut the tension strap to length pipe circumfere...

Страница 51: ...the pipe if possible mount the metal spring if necessary opposite to the tension strap clamp Insert the long end of the tension strap into the second slot of the tension strap clamp see Fig 6 42 a Tig...

Страница 52: ...Place the tension strap around the pipe see Fig 6 46 Position the ratchet clasp tension strap clamp and metal spring if necessary on the pipe on a horizontal pipe mount the tension strap clamp on the...

Страница 53: ...screws Notice In order to release the screw and the tension strap press the lever down see Fig 6 47 Fig 6 46 Ratchet clasp with tension strap tension strap clamp and metal spring on the pipe 1 metal...

Страница 54: ...sioning screw has to be placed above the hole in the transducer see Fig 6 49 and Fig 6 50 Correct the transducer distance if necessary Tighten the tensioning screws Fix the spacing element on the rail...

Страница 55: ...ove the cover Insert the lever tool in one of the 4 openings of the cover see Fig 6 52 Press the lever tool against the fixture Bend the cover outwards and release it from the anchoring Repeat the ste...

Страница 56: ...can be installed on the pipe using screws or tension straps Select the mount ing instruction On pipes with a diameter of max 40 mm it is recommended to use screws for the installation of the PermaFiX...

Страница 57: ...the rail with tension straps The number and arrangement of PermaFiX rails depend on the required transducer arrangement and transducer distance See Fig 6 56 for an overview of possible arrangements S...

Страница 58: ...place the tension strap around the pipe Insert the loosen end of the tension strap into the second slot of the tension strap screw see Fig 6 58 Tighten the tension strap Cut off the protruding tensio...

Страница 59: ...ces the conduit sealing fitting Align the transducer with the cover in such way that the hole in the transducer is placed below the tensioning screw Use the tensioning screw of the cover to press the...

Страница 60: ...rangement with screws 1 PermaFiX 2 screws 3 conduit sealing fitting 4 flexible conduit approved for FM Class I Division 1 Fig 6 62 PermaFiX on the pipe example reflection arrangement with tension stra...

Страница 61: ...ay result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Installation connection a...

Страница 62: ...ction of the outputs and of the power supply on the rear side of the upper housing see Fig 7 1 Important Do not use objects to open the housing which may damage the thread or housing gasket Fig 7 1 Co...

Страница 63: ...n the cable gland of the transducer cable The compression part remains in the cap nut Push the transducer cable through the cap nut and the compression part Prepare the transducer cable Shorten the ex...

Страница 64: ...gland Connect the transducer cable to the terminals of the transmitter see Fig 7 4 and Tab 7 2 Notice For good electromagnetic compatibility EMC it is important to ensure good electrical contact betw...

Страница 65: ...g Fix the cable gland by screwing the cap nut onto the basic part Connect the extension cable to the terminals of the transmitter see Fig 7 5 and Tab 7 3 Tab 7 2 Terminal assignment terminal connectio...

Страница 66: ...ompression part Screw the gasket ring side of the basic part into the junction box Insert the transducer cable into the junction box Fix the cable gland by screwing the cap nut onto the basic part Con...

Страница 67: ...ucer cable by tightening the cable gland Connect the transducer cable to the terminals of the junction box see Fig 7 7 and Tab 7 5 Fig 7 6 Connection of the transducer cable with plastic cable jacket...

Страница 68: ...on cable Open the cable gland of the extension cable The compression part remains in the cap nut Push the extension cable through the cap nut and the compression part Prepare the extension cable Short...

Страница 69: ...cable internal shield TRS brown cable internal shield TR brown cable core cable gland external shield Fig 7 8 Connection of the extension and transducer cable to the junction box 1 extension cable 2...

Страница 70: ...xtension cable into the junction box Prepare the extension cable Cut the outer shield and brush it back Pull the extension cable back until the back brushed external shield is below the shield termina...

Страница 71: ...ng Connect the transducer cable to the terminals of the transmitter see Fig 7 10 and Tab 7 2 Fig 7 9 Connection of the extension and transducer cable to the junction box 1 extension cable 2 external s...

Страница 72: ...repare the extension cable Connect the extension cable to the terminals of the transmitter see Fig 7 11 and Tab 7 9 Fig 7 10 Connection of the transducer cable to the transmitter 1 transducer cable Ta...

Страница 73: ...als of the junction box see Tab 7 10 Tab 7 11 and Fig 7 12 Tab 7 10 Terminal assignment transducer cable terminal connection KL1 terminal connection KL2 V transducer core R transducer core VS transduc...

Страница 74: ...the sockets see Fig 7 13 Socket A is assigned to the transducers of measuring channel A socket B to the transducers of measuring channel B Connect the transmitter to the power supply Enter all paramet...

Страница 75: ...mitter see Fig 7 14 The external protective earth is connected to the equipotential bonding terminals of the upper and lower housing of the transmitter Connect the power cable to the transmitter For t...

Страница 76: ...he transmitter Screw the sealing ring side of the basic part into the transmitter housing Fix the cable gland by screwing the cap nut onto the basic part Connect the cable to the terminals of the tran...

Страница 77: ...F1 current output Rext 500 HART Uint 24 V passive current output HART G809 A1 G809 F1 current output Uext 7 30 V Uext 0 022 A Rext 7 V example Uext 12 V Rext 227 current during transmitter fault Ifaul...

Страница 78: ...equency output open collector NAMUR G809 A1 G809 F1 Uext 8 2 V Rc 1 k DIN EN 60947 5 6 NAMUR frequency output open collector G809 A1 G809 F1 Uext 5 24 V Rc k Uext Ic mA Ic 1 4 mA Tab 7 13 Circuit of t...

Страница 79: ...00 mA Rc k Uext Imax mA Rint circuit 2 RS485 G809 A1 G809 F1 120 termination resistor Tab 7 13 Circuit of the outputs output transmitter external circuit remark internal circuit connection The number...

Страница 80: ...serial interface of the PC If the RS232 cable cannot be connected to the PC use the RS232 USB adapter The RS232 USB adapter and the RS232 cable are part of the serial data kit optional 7 4 2 RS485 int...

Страница 81: ...result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Installation connection and...

Страница 82: ...l Press ENTER Select yes if the transmitter is to be used in the region of Canada Press ENTER This display will only be indicated if imperial is selected The current time is displayed Press ENTER to c...

Страница 83: ...SYSTEM settings contains the following menu items dialogs and menus measurement outputs storing snap network serial transmission miscellaneous clock settings libraries Tab 8 1 Operation state indicati...

Страница 84: ...s been entered the main menu is displayed in the selected language The selected language remains activated when the transmitter is switched off and on again After an initialization of the transmitter...

Страница 85: ...em at the current settings Press ENTER If yes is selected the message FACTORY DEFAULT DONE will be displayed Select yes to delete the stored measured values or no to keep them Press ENTER This display...

Страница 86: ...e in the lower line Press ENTER 8 9 Interruption of the power supply As soon as the measurement begins all current measuring parameters will be stored in a non volatile INIT resistant EPROM In case of...

Страница 87: ...may result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Installation connection...

Страница 88: ...to enter the pipe circumference instead of the outer pipe diameter see section 18 1 If the input of the pipe circumference is activated and zero is entered in Outer Diameter the menu item Pipe Cir cum...

Страница 89: ...is selected For the sound speed of some materials see annex C Lining Select yes if the pipe has a lining Select no if the pipe has no lining Press ENTER Lining material Select the lining material Pres...

Страница 90: ...ughness cannot be exactly determined and must therefore be estimated For the roughness of some materials see annex C Enter the roughness for the selected pipe or lining material Change the value accor...

Страница 91: ...nge of possible values for the sound speed has to be entered Enter the average sound speed of the fluid Press ENTER This display will only be indicated if Other Medium is selected Sound speed range of...

Страница 92: ...ty of the fluid The value entered here is used for the calculation if the fluid temperature is not measured Enter the fluid temperature The value has to be within the operating temperature of the tran...

Страница 93: ...lying the flow velocity by the cross sectional pipe area standard volumetric flow rate is calculated from the oper ating volumetric flow rate mass flow rate is calculated by multiplying the volumetric...

Страница 94: ...umetric flow rate the standard volumetric flow rate will be totalized The unit of measurement of the standard volumetric flow rate is displayed without any prefix 9 2 2 Input of the damping factor Eac...

Страница 95: ...ss ENTER The scroll list contains all available outputs of the transmitter Current Ix Binary Bx Frequency Fx A tick after a list item indicates that this output has already been installed Select yes t...

Страница 96: ...itive flow direction actual measure totalizer for the physical quantity selected in the program branch Output Options Flow flow totalizer Q totalizer for the negative flow direction actual measure tot...

Страница 97: ...be measured e g if there are solids in the fluid Select a list item for the error output see Tab 9 2 Press ENTER If Other value is selected enter an error value The value has to be within the output...

Страница 98: ...section 9 2 5 and Tab 9 3 The volumetric flow rate cannot be measured during the time interval t0 t1 The error value will be output Tab 9 3 Examples for the error output output range 4 20 mA list ite...

Страница 99: ...e menu item SYSTEM settings Press ENTER 9 2 4 Activation of an analog output Select the channel for which an output is to be activated in the program branch Output Options Press ENTER This display wil...

Страница 100: ...ent of the source item will be displayed Full Scale Val is the value assigned to the value Output MAX of the output range Press ENTER Function test The function of the output can now be tested Connect...

Страница 101: ...ansmitter has only one measuring channel Select a channel with key Press key to activate or deactivate the channel Press ENTER A deactivated channel will be ignored during the measurement The paramete...

Страница 102: ...ment arrangement The profile correction 2 0 at non ideal inflow conditions will be used Transverse flow effects will be compensated Select yes if there is only one transducer pair at the measuring poi...

Страница 103: ...transit time of the measuring signal in s S signal amplitude Q signal quality bar graph has to have max length If the signal is not sufficient for a measurement Q UNDEF will be displayed In case of l...

Страница 104: ...e which had been calculated on the basis of the sound speed range entered in the program branch Parameter If the difference between the optimum and entered transducer distance is less than specified i...

Страница 105: ...channels The flow rate is measured on one measuring channel for approx 1 s then the multiplexer switches to the next activated channel The time necessary for the measurement depends on the measuring c...

Страница 106: ...ayed in the lower line Press key during the measurement to change the display in the upper line press key to change the display in the lower line The character indicates that the displayed value here...

Страница 107: ...xpected sound speed of the fluid The expected sound speed is calculated from the fluid parameters OK is equal to the expected value 20 of the expected value 20 of the expected value unknown cannot be...

Страница 108: ...p of the measurement The measurement is interrupted by pressing key BRK if it is not protected by a program code see section 18 6 Tab 9 6 Commands that can be executed during the measurement command e...

Страница 109: ...Division 1 This may result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Service...

Страница 110: ...displayed in the lower line Track down the situations when the error is displayed Contact FLEXIM The transmitter does not react when key BRK is pressed during the measurement The program code is activ...

Страница 111: ...shed The measurement can be interrupted by a temporarily high portion of liquid and solids in the fluid The measured values substantially differ from the expected values Wrong measured values are ofte...

Страница 112: ...zero All derived quantities are also set to zero The cut off flow has to be set to a low value to be able to measure at low flow velocities default 2 5 cm s The entered pipe roughness is not appropria...

Страница 113: ...M Class I Division 1 This may result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warnin...

Страница 114: ...measuring point during installation Wear the required personal protective equipment Observe the applicable rules item maintenance step interval measure stainless steel housing junction box transducer...

Страница 115: ...ed 11 4 Examination of the O ring The O rings used in the transmitter see Fig 11 1 have to be regularly checked see section 11 1 as well as after each opening of the housing in order to detect possibl...

Страница 116: ...ss must be regularly checked for damages Cracked or scratched sight glasses or damaged housings have to be replaced by FLEXIM personnel Danger Risk of explosion when using the transmitter in explosive...

Страница 117: ...ivision 1 This may result in personal or material damage or other dangerous situations Observe the Safety instructions for the use in explosive atmospheres see document SIFLUXUS_1N62 Warning Installat...

Страница 118: ...ing is carried out in reverse order to its installation see chapter 6 12 2 Disposal The measuring equipment has to be disposed in accordance to the applicable regulations Depending on the material the...

Страница 119: ...nstall or reconfigure the output Press ENTER Select no to uninstall the output and to return to the previous menu item in order to select another output Press ENTER Assignment of a measuring channel S...

Страница 120: ...volumetric flow rate The frequency output can be configured in such way that the source item can be totalized by using each period of the output signal as the increment Installation of a frequency out...

Страница 121: ...Select the channel for which the output is to be activated in the program branch Output Options Press ENTER This display will not be indicated if the transmitter has only one measuring channel This di...

Страница 122: ...in this case because the integration of the flow velocity does not result in a rea sonable value Enter the pulse value The unit of measurement will be displayed according to the actual physical quanti...

Страница 123: ...the channel for which the output is to be activated in the program branch Output Options Press ENTER This display will not be indicated if the transmitter has only one measuring channel Press ENTER u...

Страница 124: ...r is deactivated the storing of measured values will be stopped Select ON to activate the ringbuffer Press ENTER Storage mode Select the storage mode Press ENTER If sample is selected the current meas...

Страница 125: ...gether with the measured values Press ENTER Option of flow velocity storing Enter the HotCode 007043 Select auto to store the flow velocity as an integer Select full to store the flow velocity as a fl...

Страница 126: ...a is activated and Special Funct SYSTEM settings Ringbuffer is de activated a message indicating an error will be displayed as soon as the data logger is full Press ENTER The error message will be dis...

Страница 127: ...ber of the transmitter are displayed in the upper line The available data logger will be displayed in the lower line here 18 327 additional measured values can be stored Press key BRK twice to return...

Страница 128: ...Online transmission The measured values are transmitted during the measurement The data logger works independently of the online transmission but with the same transmission rate Start the terminal pro...

Страница 129: ...S485 interface Transmitter settings Select Special Funct Print Meas Val Press ENTER The following message will be displayed if no measured values are stored Press ENTER The following message will be d...

Страница 130: ...t Special Funct SYSTEM settings Network to change the settings of the transmission parameters Press ENTER This display will only be indicated if the transmitter has an RS485 interface Press ENTER to c...

Страница 131: ...e transmitter and the measurement The following lines contain the parameters for each channel Next the line DATA is transmitted Followed by the column titles see Tab 15 2 for the corresponding channel...

Страница 132: ...generated Offline transmission During the offline transmission of data columns will only be created if at least one measured value is stored in the data set The columns Q_POS and Q_NEG will not be ge...

Страница 133: ...automatic toggling of the totalizer display between positive and negative flow direction can be set Enter a time interval between 0 off and 5 s Press ENTER Totalizer storing It is possible to store t...

Страница 134: ...r limit These measured values will be marked as outliers The upper limit of the flow velocity is set in Special Funct SYSTEM settings Measuring Velocity limit Enter zero to switch off the outliers che...

Страница 135: ...he absolute value of the flow velocity Press ENTER Select factory to use the default limit of 2 5 cm s 0 025 m s for the cut off flow Select user to enter the cut off flow Press ENTER If Cut off Flow...

Страница 136: ...ect a list item default on Press ENTER Input of the disturbance distance If With disturbance is selected in the menu item Special Funct SYSTEM settings Measuring ProfileCorr 2 0 the disturbance distan...

Страница 137: ...e is used It is not possible to change the measuring channel The measurement takes place on one channel only As long as the FastFood mode is activated no measurement is carried out on the other channe...

Страница 138: ...ured values of a measuring channel All functions which are possible with the measured values of a measuring channel totalizing online transmission of data storing output etc can also be done with the...

Страница 139: ...is selected The measured values of the calculation channel here Y will be valid if the measured values of both measuring channels here A and B are valid If only one measuring channel provides valid m...

Страница 140: ...o activate the data logger and no to deactivate it Press ENTER Tab 16 1 Physical quantities of the calculation channels physical quantity of the calculation channel possible physical quantity of the f...

Страница 141: ...cted the measured values of the measuring channels will be displayed alternately but not the measured values of the calculation channels Press key to display the calculation function Press key to disp...

Страница 142: ...cer type Enter the HotCode 071001 see section 8 4 Enter the lower limit of the inner pipe diameter of the displayed transducer type Press ENTER to select the next transducer type 16 10 Transducer temp...

Страница 143: ...s Notice The menu item Alarm Output will only be displayed in the program branch Output Options if a binary output has been installed see section 13 1 Output Options for Channel A Tab 16 2 Alarm prope...

Страница 144: ...alarm will switch if the measured value falls below the limit If the switching condition QUANT is selected in the scroll list func the limit of the output has to be defined Enter the limit of the tot...

Страница 145: ...s This might cause an apparent switching delay when the measured value changes marginally less than 2 decimal places In this case the switching accuracy of the output is higher than the accu racy of t...

Страница 146: ...rm remains activated for at least 1 s even if the switching condition is met for a shorter period of time Alarm outputs with the switching condition QUANT will be activated if the limit is reached Ala...

Страница 147: ...rm output indication is activated the alarm state can be displayed during the measurement Press key to scroll through the upper line or key in the lower line until the alarm state is displayed 16 12 5...

Страница 148: ...of sound paths has to be entered Some menu items that are not visible in the normal mode are displayed 17 1 Activation deactivation Enter the HotCode 071049 see section 8 4 Press ENTER It is displayed...

Страница 149: ...Press ENTER Select the measuring channel for which the flow parameters are to be defined here Channel A Press ENTER 17 3 1 Profile bounds Select user to define the profile bounds If factory is selecte...

Страница 150: ...o the menu item SYSTEM settings If on is selected enter the slope The input of zero deactivates the correction Press ENTER Enter the offset Enter zero to work without offset Press ENTER Special Funct...

Страница 151: ...current value of the amplification GAIN can be displayed in the upper line in the program branch Measuring If the current value of the amplification is higher than the max amplification the current v...

Страница 152: ...for each measuring channel the offset Enter zero to use the default value of 300 m s Press ENTER It is possible to display the valid range for the sound speed SS in the lower line within the program...

Страница 153: ...ENTER until the menu item Total digits is displayed Select one of the following list items Automatic dynamic adaptation Fixed to x digit x decimal places range 0 4 Press ENTER Total digits Automatic T...

Страница 154: ...vate the display of the totalizer sum off to deactivate it Press ENTER If the display of the totalizer sum is activated the sum Q of the totalizers can be displayed in the upper line during the measur...

Страница 155: ...owing parameters can be displayed during the measurement in the SuperUser mode display explanation t transit time of the measuring signal in the fluid c sound speed REYNOLD Reynolds number VARI A stan...

Страница 156: ...erence are calculated on the basis of the limits for the outer pipe diameter Press ENTER During the next scroll through the program branch Parameter the outer pipe diameter that corresponds to the ent...

Страница 157: ...lowed by the entered transducer distance if the recommended and the entered transducer distance are not identical During the transducer positioning in the program branch Measuring only the entered tra...

Страница 158: ...the absolute or the relative pressure is to be used Select on or off Press ENTER If on is selected the absolute pressure pa will be displayed input transmitted If off is selected the relative pressure...

Страница 159: ...0 on profile correction 2 0 at ideal inflow conditions default With disturbance profile correction 2 0 at non ideal inflow conditions Press ENTER Select normal to display and transmit the profile cor...

Страница 160: ...w velocity greater than this limit will be set to zero If Cut off Flow absolut and user are selected only one value have to be entered Enter the cut off flow Press ENTER The absolute value of all flow...

Страница 161: ...terial list to edit the material scroll list or Medium list to edit the fluid scroll list Select go back to return to the menu item SYSTEM settings Press ENTER Select factory if all materials fluids o...

Страница 162: ...Add all to add all materials fluids of the database to the current scroll list Press ENTER 18 4 4 Removing a material fluid from the scroll list Select Remove Material or Remove Medium to remove a ma...

Страница 163: ...transmitter The display contrast is adjusted with the following keys Press ENTER It is possible to reset the display to medium contrast by means of a HotCode Enter the HotCode 555000 see section 8 4 N...

Страница 164: ...long as no other valid program code is entered or the program code is not deactivated 18 6 2 Intervention in the measurement If a program code is activated the message PROGRAM CODE ACTIVE will be disp...

Страница 165: ...the program code with the keys and Press ENTER At first 000000 is displayed If the program code starts with 000 ENTER can be pressed immediately 18 6 3 Deactivation of the program code Select Special...

Страница 166: ...18 Settings 18 6 Program code FLUXUS G809 2020 06 25 UMFLUXUS_G809V2 2EN 166...

Страница 167: ...ing channel System settings Pipe parameters Measuring point number Physical quantity Dialogs and menus Measurement Outputs Storing Serial transmission Miscellaneous Set Slock Libraries Fluid parameter...

Страница 168: ...terial Lining yes no Lining Liner Thickness Roughness Disturb distance 1 Medium Medium Temperat Fluid pressure Transducer Type Special Version Standard Transd Data 1 Transd Data 2 Transd Data 3 Transd...

Страница 169: ...Annex FLUXUS G809 A Menu structure 169 UMFLUXUS_G809V2 2EN 2020 06 25 Measurement settings see chapter 9 Main menu Output Options for Channel x Physic Quant Unit of measurement Damping...

Страница 170: ...Output Options SYSTEM settings for Channel x ENTER until Storing Store Meas Data Start logger Ringbuffer yes no Storage Rate yes no Storage mode sample average Quantity Storage Main menu one both Spe...

Страница 171: ...chapter 9 Main menu Measuring Channel Meas Point No PROFILE CORR 1 Sound Path Transd Distance Signal test Transd Distance Measurement display BRK Measurement stop Legend 1 this will only be displayed...

Страница 172: ...9 Main menu Main menu Special Funct Output Options for Channel x SYSTEM settings ENTER until Proc outputs Current Loop Install Output Current Frequency Binary yes no Zero Scale Val Enable Full Scale V...

Страница 173: ...SYSTEM settings Calibration Measuring for Channel x Miscellaneous ENTER until factory user Gain threshold Laminar flow Bad soundspeed Turbulent flow Thresh Offset Calibration ENTER until off on Do not...

Страница 174: ...tput Options SYSTEM settings for Channel x Proc outputs ENTER until Install Output Alarm Output Binary Enable func typ mode Source chan MAX MIN QUANT ERROR OFF NON HOLD HOLD NO Cont NC Cont Source ite...

Страница 175: ...enheit unit of measurement description bar a bar absolute bar g bar relative psi a pound per square inch absolute psi g pound per square inch relative unit of measurement description g cm gram per cub...

Страница 176: ...egaliter d megaliter per day Ml bbl d barrel per day bbl bbl h barrel per hour bbl bbl m barrel per minute bbl bbl s barrel per second bbl USgpd US gal d gallon per day gal USgph US gal h gallon per h...

Страница 177: ...lized t h ton per hour t t d ton per day t kg h kilogram per hour kg kg min kilogram per minute kg kg s kilogram per second kg g s gram per second g lb d pound per day lb lb h pound per hour lb lb m p...

Страница 178: ...tion ctrans m s clong m s cflow Carbon Steel steel normal 3230 5930 trans Stainless Steel steel stainless 3100 5790 trans DUPLEX duplex stainless steel 3272 5720 trans Ductile Iron ductile iron 2650 t...

Страница 179: ...ing 0 12 new without lining 0 25 1 rusted 1 1 5 encrusted 1 5 3 fluid display explanation sound speed m s kinematic viscosity mm2 s density g cm3 Aceton acetone 1190 0 4 0 7300 Ammonia NH3 ammonia NH3...

Страница 180: ...248 Paraffin 248 1468 195 1 0 8450 R134 Freon R134 Freon 522 0 2 1 2400 R22 Freon R22 Freon 558 0 1 1 2130 Crudeoil hi API crude oil light 1163 14 0 0 8130 Crudeoil low API crude oil heavy 1370 639 5...

Страница 181: ...22 583 481 54 0 595270779 0 9677784 0 80 68 928 411 41 0 184177693 0 819764 10 64 534 422 6 0 19880993 0 8446627 20 60 824 433 08 0 213649217 0 8656106 30 57 632 442 93 0 228709745 0 883441 40 54 841...

Отзывы: