6 Installation
FLUXUS G809
6.2 Transducers
59
UMFLUXUS_G809V2-2EN, 2020-06-25
Installation of the transducers
Mount the transducers into the rail in such way that the engravings on the transducers form an arrow , see Fig. 6.60. The
transducer cables show in opposite directions.
• Insert the transducer with the conduit sealing fitting into the rail.
• Observe that there must be no air pockets between the transducer contact surface and the pipe wall.
• Slide the cover over the rail. The round cutout of the cover faces the conduit sealing fitting.
• Align the transducer with the cover in such way that the hole in the transducer is placed below the tensioning screw. Use
the tensioning screw of the cover to press the transducer against the pipe.
• Screw the cable conduit (approved for FM Class I, Division 1) to the conduit sealing fitting of the transducer.
• Repeat the steps for the second transducer.
• Connect the flexible conduit to the conduit system of the facility in accordance with the local regulations.
Fig. 6.59: Installation of the rail with tension straps (diagonal arrangement, small transducer distance)
1 – rail
2 – tension straps
Fig. 6.60: Correct positioning of the transducers
1 – holes for the screws
2 – engravings
1
2
1
2