15 Data transmission
FLUXUS G809
15.2 Terminal program
129
UMFLUXUS_G809V2-2EN, 2020-06-25
15.2.2
Offline transmission
• Start the terminal program.
• Enter the transmission parameters into the terminal program. The transmission parameters of the terminal program and
the transmitter have to be identical, see section 15.3.
Selection of the interface for the offline transmission of data
• Select
Special Funct.\SYSTEM settings\serial transmis.
• Press ENTER until
Send Offline via
is displayed.
• Select the serial interface for the offline transmission of data.
This display will only be indicated if the transmitter has an RS485 interface.
Transmitter settings
• Select
Special Funct.\Print Meas.Val.
• Press ENTER.
The following message will be displayed if no measured values are stored.
• Press ENTER.
The following message will be displayed if the measurement values are transmitted.
The progress of the transmission of data is displayed by a bar graph.
The following error message will be displayed if an error has occurred during the serial transmission.
• Press ENTER.
• Check the connections and make sure the PC is ready to receive data.
Notice!
During the offline transmission only those data is transmitted that is stored in the transmitter.
Special Funct.\SYSTEM settings\serial transmis.
Special Funct.\SYSTEM settings\serial transmis.\Send Offline via
Special Funct.\Print Meas.Val.
NO VALUES
Print Meas.Val.
Send Header
.................
■■■■■■
.................
SERIAL ERROR
Print Meas.Val.