8 Start-up
FLUXUS G809
8.6 Initialization
85
UMFLUXUS_G809V2-2EN, 2020-06-25
8.6
Initialization
During an initialization (INIT) of the transmitter, the settings in the program branches
Parameter
and
Output Options
and some of the settings in the program branch
Special Funct.
are reset to the default settings of the manufacturer.
Proceed as follows to execute an initialization:
• When switching on the transmitter: keep keys BRK and CLR pressed.
• During the operation of the transmitter: press keys BRK, CLR and ENTER at the same time. A reset is executed. Release
only key ENTER. Keep keys BRK and CLR pressed.
After the initialization has been executed, the message
INITIALISATION DONE
is displayed.
After the initialization, the remaining settings of the transmitter can be reset to the default settings and/or the stored mea
-
sured values can be deleted.
• Select
yes
to reset the remaining settings of the transmitter to default settings or
no
to keep them at the current settings.
• Press ENTER.
If
yes
is selected, the message
FACTORY DEFAULT
DONE
will be displayed.
• Select
yes
to delete the stored measured values or
no
to keep them.
• Press ENTER.
This display will only be indicated if measured values are stored in the transmitter.
8.7
Time and date
The transmitter has a battery-powered clock. Measured values are automatically stored with date and time.
• Select the menu item
Set Clock
.
• Press ENTER.
The current time is displayed.
• Select
ok
to confirm the time or
new
to adjust it.
• Press ENTER.
• Select the character to be edited with key
. Edit the selected character with key
.
• Press ENTER.
The new time is displayed.
• Select
ok
to confirm the time or
new
to adjust it again.
• Press ENTER.
After the time has been set, the date is displayed.
• Select
ok
to confirm the date or
new
to adjust it.
• Press ENTER.
• Select the character to be edited with key
. Edit the selected character with key
.
• Press ENTER.
The new date is displayed.
• Select
ok
to confirm the date or
new
to adjust it again.
• Press ENTER.
FACTORY DEFAULT
Delete Meas.Val.
Special Funct.\SYSTEM settings\Set Clock\TIME
Special Funct.\SYSTEM settings\Set Clock\DATE