38
This value depends on the
cropping configuration and the
number of bits per pixel.
Image correction
The image corrections that can be applied to the frames in real time are the following.
Table 24 - Description of the image corrections available
Name
Access Value
Description
GlowReduction
RW
Off / On
Glow reduction is disabled / applied.
Trigger mode
The acquisition can be triggered by various signals. The registers used to configure this mode are the
following.
Name
Access
Values
Description
TriggerSelector
RW
AcquisitionStart
Trigger will be set up to start a continuous
acquisition.
AcquisitionEnd
Trigger will be set up to stop a continuous
acquisition.
FrameBurstStart Trigger will be set up to start an
acquisition of
“AcquisitionBurstFrameCount” frames.
FrameBurstEnd
Trigger will be set up to stop an
acquisition of
“AcquisitionBurstFrameCount” frames.
FrameStart
Trigger will be set up to start a single
frame acquisition.
FrameEnd
Not implemented
ExposureStart
Trigger will be set up to start the exposure
of a single frame acquisition.
ExposureEnd
Trigger will be set up to stop the exposure
of a single frame acquisition.
TriggerMode
[TriggerSelector]
RW
ON
The selected trigger is enabled
OFF
The selected trigger is disabled
TriggerSource
[TriggerSelector]
RW
Software
The selected trigger will be software
generated.
Line0
The selected trigger will be sent through
the LEMO 0 I/0 connector.
Line1
The selected trigger will be sent through
the LEMO 1 I/0 connector.
LinkTrigger0
The selected trigger will be sent through
the master Coaxpress cable.
LinkTrigger1
The selected trigger will be sent through
the slave Coaxpress cable.
Содержание C-BLUE One