36
Basic device control
Table 20 – Description of the commands to control the camera
Name
Access
Values
Description
DeviceReset
WO
Command
Resets the device. After reset, the device must
be rediscovered.
DeviceShutdown
WO
Command
Turns the device off.
Temperature control
The temperature of the camera can be monitored thanks to several measurement probes.
Table 21 - Description of the SFNC registers associated to temperature control
Name
Acces
s
Value
Description
DeviceTemperatureSele
ctor
RW
Sensor, CPU,
Power, Frontend,
Heatsink, Case
Temperature of the defined part
DeviceTemperature
RO
Device temperature in degrees Celsius.
It is measured at the location selected
by
DeviceTemperatureSelector
.
DeviceTecSelector
RW
TEC1
Primary TEC
DeviceTecVoltage
RO
Float
Voltage applied to TEC in volts.
Measured at the TEC selected using
DeviceTecSelector.
DeviceTecCurrent
RO
Float
Current consumed by the TEC in
Amperes. Measured at the TEC selected
using DeviceTecSelector.
DeviceTecPower
RO
Float
TEC power consumption in Watts.
Measured at the TEC selected using
DeviceTecSelector.
DeviceFanMode
RW
Automatic
Manual
The fan speed is controlled by the user
or automatically.
DeviceFanSpeed
RW
Integer
Selects the speed of the fan in manual
mode.
DeviceCoolingEnable
RW
Boolean
Controls if the sensor cooling is enabled.
DeviceCoolingSetpoint
RW
Float
Specifies the sensor temperature target
when cooling is enabled
Acquisition Control
Basic commands
The acquisition mode and the commands available to control the acquisition are the following.