Table 37 - Common issues and troubleshooting procedures
Issue
Possible cause
Troubleshooting
Cooling does not reach
the set point
The cooling plate is not in good thermal contact.
The cooling system is off or defective.
Check in Table 9 that the set points are within the reach of the system.
Make sure the camera’s fan is not blocked.
Make sure the cooling system is switched on, filled with liquid fluid and
that the liquid is flowing.
The camera has shut
down automatically
The camera has overheated.
The external power supply has been plugged in
when the camera was powered in PoCXP
Reboot the camera by unplugging/ replugging the power cord.
No signal on the status
LED
The camera is not powered
The LEDs may have been turned off
Check that the camera is correctly powered
Make sure the LEDs have not been turned off
The camera does not
make images
The camera is not properly connected
The CoaXPress® cable are inverted
..
Check that:
Cables are connected properly.
CoaXPress® cables are in good position.
Graphical User Interface software is running properly
Camera responds to software commands (i.e. thermal reading,
monitoring functions)
Then:
Restart the camera
Restart the computer
The camera is not
detected
The CoaXPress® cable are not in a correct
position.
Check that the camera is powered properly.
Check the CoaXPress® connectors are well connected.
Invert the CoaXPress® cables.
No power up with PoCXP
connection
The grabber may be configured to disable
automatic PoCXP connection.
Check that PoCXP discovery is enabled.