34
10.1.3 Employ the last unit weight when power on
To employ this function, set F35 = On. Under this setting the last unit weight (before powered off) will be
employed automatically when powered on again.
To disable this function, set F35 to Off.
Note:
- Unit weight obtained by this function is not supported by Auto Unit Weight Enhancement Function
10.1.4 Recall unit weigh unit from quick PLU
Refer to
paragraph
11.1.2
for details.
10.2 Start Counting Process
Once unit weight is entered (by either method listed on
paragraph 10.1
, instrument is ready for counting
operation. By adding more to or removing some from weighing platform, the corresponding quantity will be
displayed on LCD1 automatically.
10.3 Auto Unit Weight Enhancement Function
In order to obtain the best counting result and minimize sampling error, this instrument is equipped with Auto
Unit Weight Enhancement Function. This function will automatically be employed when unit weight is
determined by sampling method
7
.
Auto Unit Weight Enhancement Function starts automatically when
all
of the below are met: -
a.
Unit weight is obtained by sampling method as listed on
10.1.2
.
b.
6 pieces
≦
quantity
added
≦
9999 pieces.
c.
6 pieces
≦
quantity
added
≦
100% of last total quantity on weighing platform.
Note
: - Auto Unit Weight Enhancement Function stops automatically when total quantity on weighing
platform reaches 10000 pieces or more.
When all the above requirements are met, a new unit weight will be calculated and displayed on LCD3
automatically.
10.4 Check Function
8
9
10
Check function is used to compare current weighing or counting result with the preset Lo and Hi Limit. The
comparison results (Lo, OK, Hi) will then be displayed in different backlight colors with or without buzzer
11
.
Check results are also sent to Control Output Port
12
.
Check functions accept various values value and targets for P1 and P2. Settings for P1 and P2 are
independent to each other.
7
Refer to
paragraph 10.1.2
for details.
8
Check mode will not operate when weight is less than 20d (or 20d1 for dual weighing range/interval mode).
9
Set also F15 for desired Check buzzer output.
10
If necessary, set F26 to meeting operation requirements.
11
Set F15 for preferred buzzer output configuration.
12
Refer to
paragraph 4.3
for details.