31
9.4 Warm Up Time
Allow a warm up period of not less than 60 seconds after powered on before using.
9.5 Dual Platform Operation
If the secondary platform is ordered but it is not enabled, contact your dealer for support.
9.5.1 Select preferred weighing platform by [PF/Set]
To enable platform selection through the
[PF/Set]
, set internal function F31 to
P.F.
After that, by pressing
[PF/Set]
, instrument will shift among P1 only, P2 only and both. Platform (s) in operation is indicated by
P1
or
P2
or
PA
Indicator.
9.5.2 Summing up transaction results of both platforms
Transaction results of P1 and P2 can be totalized (sum total). To enable platform selection through the
[PF/Set]
,
set internal function F9 to
On
.
•
When only
P1 Indicator
appears = Only P1 is in operation.
•
When only
P2 Indicator
appears = Only P2 is in operation.
•
When both P1 and P2 indicators appear = Both P1 and P2 (hereinafter P1 + P2 is referred as PA) are
in operation and transaction results of both platforms are totalized and being displayed.
9.6 Keyboard Lock
When keyboard lock is enabled in F20, only
[On/Off]
,
[Zero]
,
[Tare]
and
[On/Off]
keys can function during
normal operation.
9.7 Loads and Weighing Platform
•
Always place an object onto platform gently. Excessive force / shock applied to platform may cause un-
recoverable damage to the weight sensor inside platform.
•
It is a good practice to remove all loads from platform after weighing. It will prolong the life of the weight
sensor.
9.8 Manual Zero
If zero weight cannot be obtained when unloaded, press
[Zero]
.
Zero Indicator
lights up when instrument is at
zero status.
9.9 Tare Functions
9.9.1 Manual and automatic tare
This instrument is equipped with the below auto tare functions. These functions can be selected in F12.
a.
Off: - Manual Tare
b.
Auto: - Instrument will assume the first stable weight (≥ 20d or 20d
1
) applied is a container and will then
tare off the weight of it automatically. When container is removed and gross weight result is zero, tare
effect will be cancelled automatically.
c.
Contin (continuous auto tare): - All stable weight (≥ 20d or 20d
1
) applied will be tared off automatically.
When all loads are removed and gross weight result is zero, tare effect will be cancelled automatically.
Notes: -
•
If Contin is set, select also d.t. (delay time, 0.0 ~ 9.9 second
.
Default = 1.5 second
).
•
Delay time is the time duration from when a stable weight is detected until it is automatically tared off.
•
Enter the preferred d.t. value through numeric keys and then press
[Print/M+]
to save.
Содержание DC-10