No: - Instrument will not ask for operation number when instrument power on.
Notes: -
•
If a different operator number (than what is running now) is entered, instrument displays
Clear
.
•
The operator number entered will be carried out automatically on output formats Lab 3
~ Lab 5.
F29
Read Calibration & Parameter Counts and Pointer Value
•
C (Calibration count): - shows total times of calibration.
•
O (Parameter set count): - shows total times that the important parameters (F80~F88)
has been altered.
Press
[Zero]
to quit to F29 or
[Check]
to display the current pointer value. Pointer Value is the
system serial number of the next transaction to be saved to Alibi memory.
F30
Allow Letters and
Symbols for
Customer &
Product Code
Manual Entry
** No
Yes
F31
[PF/Set]
Key
Assignment
** Set
(Access to Internal Function
Setting when in Operation Status)
P.F.
(Shift between P1 and P2)
Note
: - When F79 is set to
Off
, Mode Key is always =
Set
and this internal function is not
accessible.
F32
External Input
Assignment
in1
(Input 1)
in2
(Input 2)
in3
(Input 3)
1 of the below 7 input parameters can be assigned to each external input.
•
ZEro =
[Zero]
•
tAre =
[Tare]
•
Print =
[Print/M+]
•
ACC-Mr =
[MR]
•
CLEAr =
[CE/x10]
•
P.F.= Shift among P1, P2 and PA.
•
Set = Access internal function setting mode (F1~F36).
Setting procedures:
-
1.
Go to F32 then press
[Print/M+]
.
2.
Name of the current external input No. is displayed on LCD2.
3.
Press
[MR]
or
[PF/Set]
until the preferred external input point No. appears on LCD2
then press
[Print/M+]
.
4.
Press
[MR]
or
[PF/Set]
until the preferred input parameter appears on LCD3 then press
[Print/M+]
to save
5.
Instrument displays the next external input point No. on LCD2.
6.
Repeat above steps
2
~
5
for all other external input points.
F33
Report Output
Comport &
Content Settings
Set report output comport No. and report content setting