Setting procedures
: -
1.
Go to F33 then press
[Print/M+]
.
2.
Name of the report output port is displayed on LCD2.
Note
: - avoid assigning RS485 for report output unless Machine ID (set in F18) is =
blank.
3.
Press
[MR]
or
[PF/Set]
until the preferred report output port appears on LCD2 then
press
[Print/M+]
.
4.
Instrument displays report content detail setting and parameter on LCD2 and LCD3
respectively. At this point, report output comport No. is set. Press
[Zero]
to quit to F33
or continue with below to set preferred report content detail parameters.
5.
Press
[MR]
or
[PF/Set]
until the preferred report content detail parameter appears on
LCD3 then press
[Print/M+]
.
Notes: -
3 report content detail parameters (Off, All and Custom) are available: -
•
If Off is selected, instrument returns to F33. In this case, output report contents
only the following data.
o
Alibi Transaction Memory Serial No.,
o
Date of the individual transaction,
o
Net Weight of the individual transaction,
o
Tare weight of individual transaction,
o
Gross weight of individual transaction.
•
If All is selected, instrument returns to F33. In this case, all contents will be
outputted.
•
If Custom is selected, follow the below procedures for report content detail
settings.
6.
Press
[MR]
or
[PF/Set]
until the preferred report content detail parameter appears on
LCD3 then press
[Print/M+]
. Below parameters are available.
•
oP.Code (Operator No.)
•
P.F. (Platform #)
•
id (Machine ID)
•
Group (Machine Group No.)
•
Time (Time of the individual transaction)
•
C Code (Customer Code)
•
C_dESC (Customer Description)
•
P Code (Product Code)
•
P_dESC (Product Description)
•
qty (No. of pieces)
•
Ref (Lo and Hi limit value of the individual transaction)
•
Result (Check result of the individual)
•
Title
•
Header
•
Footer
7.
Press
[PF/Set]
until the preferred parameter (On / Off) appears on LCD3 then press
[Print/M+]
to save and go to next report content detail parameter setting.
8.
Repeat above steps
6
and
7
to set all other report content detail until instrument returns
to F33. At this point, report content detail setting is completed.
Refer to paragraph
14
for more information.