Section 5: Troubleshooting
5.4 Troubleshooting Procedures
0101-8242-0, Rev. C
5-21
CV-6SLX User Manual
4. If the symptom is not in evidence during Step 3, troubleshoot the HV path and emitter assembly, following the instructions
in the e-gun manual. If the symptom persists during Step 3, reconnect the HV output cable to J101, switch the power
supply back on, and proceed with the troubleshooting steps described below.
5.
Fault conditions whose investigation
requires removal of HVPS top cover.
5.
Contact Temescal Field Service
.
DANGER: HIGH VOLTAGE
Before performing any of the procedures described on this page, perform the
procedure described in section 5.4.1.
Table 5-10 Filament Current Out Of Regulation
Symptom: When gun and HV are switched on, filament current ramps up and goes out of regulation
Possible Causes
Procedure
1. Faulty or loose connections.
1. Make sure that all HV and FPS connections are correctly and securely made.
2. Short in HV path
2. Check emitter, feethroughs, and cables for signs of shorts.
3. Emitter damaged
3. Follow instructions in gun manual to troubleshoot emitter issues.
4.
Fault conditions whose investigation
requires removal of FPS top cover.
4.
Contact Temescal Field Service
.
Table 5-11 HVPS ARC/OUT OF REGULATION LED Flashes Excessively
Condition indicated: Subthreshold arcing is occurring and/or HV is intermittently out of regulation
NOTE: A problem exists only if arcing is excessive for the material or process in question
Symptoms:
•
HVPS front panel: The
ARC/OUT OF REGULATION
LED flashes excessively. The
HV OUT OF REGULATION
LED turns yellow for 2 seconds each time the
ARC/OUT OF REGULATION
LED flashes
For additional information about this fault mode, see Ishikawa Diagram #6 in Appendix A.
Possible Cause
Procedure
1. Emitter damaged
1. Follow instructions in gun manual to troubleshoot emitter issues.
2. Faulty or loose connections.
2. Make sure that all HV and FPS connections are correctly and securely made.
3. HV feedthrough shorted
3. Check copper bus bars and HV feedthroughs for shorts, coatings, cracks or
arc marks
4. Chamber Pressure too high
4. Chamber pressure should be less than 1 x 10
-5
mTorr.
5. System improperly grounded for HV.
5. Follow grounding instructions provided in e-gun manual.
6. Incorrect or fluctuating line voltage
6. Confirm input line voltage
7.
Fault conditions whose investigation
requires removal of HVPS top cover.
7. Contact Temescal Field Service.
Содержание Temescal CV-6SLX
Страница 4: ...CV 6SLX User Manual ii 0101 8242 0 Rev C...
Страница 16: ...1 5 TemEBeam Controller EBC Section 1 Introduction to the CV 6SLX CV 6SLX User Manual 0101 8242 0 Rev C 1 4...
Страница 34: ...2 6 Cable Installation Section 2 Installation CV 6SLX User Manual 0101 8242 0 Rev C 2 18...
Страница 90: ...5 5 FPS Fuse Replacement Procedure Section 5 Troubleshooting CV 6SLX User Manual 5 32 0101 8242 0 Rev C...
Страница 92: ...Appendix A Pinout Table for Rear Panel Diagnostic Port J3 CV 6SLX User Manual A 2 0101 8242 0 Rev C...