5.4 Troubleshooting Procedures
Section 5: Troubleshooting
CV-6SLX User Manual
5-18
0101-8242-0, Rev. C
Table 5-4 HVPS and FPS Are Powered Up but GUN READY State Not Achieved/Confirmed
Symptoms
•
TCS displays
SRC1_GON_ALM
( = Source 1 Gun Control GO ON Timeout Alarm)
•
EBC issues
FPS Gun Ctl Go On Timeout
alarm
Possible Causes
Procedure
1. Confirm that all interlocks are made.
1.
See Table 5-2.
2. Faulty or loose cable connections.
2.
Make sure that all power module and FPS cable connections are correctly and
securely made.
3. GUN ENABLE input at Pin 6 on FPS
connector J101 is high.
3.
Check input at Pin 6. If high, proceed to Step 4. If low, investigate the GUN
ENABLE loop upstream of the FPS.
4.
Fault condition(s) whose investigation
requires removal of FPS top cover.
4.
Contact Temescal Field Service.
DANGER: HIGH VOLTAGE
Before performing the following procedure, perform the procedure described in
section 5.4.1.
Table 5-5 HVPS Powered Up but HV Does Not Switch On When Commanded To Do So
Symptoms:
•
TCS displays
PSU1_GON_ALM
(= Power Supply Control Go On Timeout Alarm)
•
HVPS Front Panel:
HV ON
LED fails to light Green
•
EBC issues
HVPS Go On Failure
alarm
Possible Causes
Procedure
1. HV READY signal is not present.
1.
See Table 5-3.
2. After remote HV GO ON command is
issued, HV ON
signals are not supplied
via Pins 8-9 of HVPS J4.
2.
Verify that the remote HV GO ON signal is supplied by EBC or PLC-based
system controller.
If the preceding steps have failed to correct the fault, perform the following procedure:
1. Switch off the power supply.
2. Disconnect HV output cable.
3. Switch on the power supply and the HV and leave them switched on for no more than 9 seconds, observing whether the
symptom persists during that interval.
CAUTION
DO NOT leave the high voltage on for more than 9 seconds with the HV cable disconnected. Doing so will
result in the emitter’s filament burning out.
4. If the symptom is not in evidence during Step 3, troubleshoot the HV path and emitter assembly, following the instructions
in the e-gun manual. If the symptom persists during Step 3, reconnect the HV output cable to J101, switch the power
supply back on, and proceed with the troubleshooting steps described below.
3.
Fault condition(s) whose investigation
requires removal of HVPS top cover.
3.
Contact Temescal Field Service
.
Содержание Temescal CV-6SLX
Страница 4: ...CV 6SLX User Manual ii 0101 8242 0 Rev C...
Страница 16: ...1 5 TemEBeam Controller EBC Section 1 Introduction to the CV 6SLX CV 6SLX User Manual 0101 8242 0 Rev C 1 4...
Страница 34: ...2 6 Cable Installation Section 2 Installation CV 6SLX User Manual 0101 8242 0 Rev C 2 18...
Страница 90: ...5 5 FPS Fuse Replacement Procedure Section 5 Troubleshooting CV 6SLX User Manual 5 32 0101 8242 0 Rev C...
Страница 92: ...Appendix A Pinout Table for Rear Panel Diagnostic Port J3 CV 6SLX User Manual A 2 0101 8242 0 Rev C...