F-960 Instruction Manual rev. 6/23/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
4
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
Carbon Dioxide (CO
₂)
SENSOR
Infrared sensor, pyroelectric detector
Nominal Range
0-100%
Full Scale Resolution
0.01% absolute
Accuracy
–
Trigger
Mode
±.5% absolute and ±3% of measured
value
Accuracy
–
Continuous
Mode
±.01% absolute and ±3% of measured
value
Lower Detection Limit
0.01%
Sampling Time
10 seconds
Zero Interval
Weekly
Calibration Interval
12 Months
Lifetime
5 years
Oxygen (O₂)
Sensor
Electrochemical
Nominal Range
0-100%
Full Scale Resolution
0.1% absolute
Accuracy
–
Trigger
Mode
±0.3% absolute and ±2% of measured
value
Accuracy
–
Continuous
Mode
±0.1% absolute and ±2% of measured
value
Lower Detection Limit
0.1%
Sampling Time
10 seconds
Zero Interval
Weekly
Calibration
6 months
Lifetime
2 years