O2 Set Span: 50% O2 standard gas
Each sensor will require its own standard gas for calibration.
After the set span is complete it is best practice to verify your
calibrations were successful by reading the standard gas to
ensure accuracy.
Settings
G.A.S. offers a settings menu allowing the user to switch
between continuous and trigger modes, as well as setting
thresholds for QA monitoring.