Connect and read each gas and record the results here after
three minutes (at least) of measure time. If any of the values are
out of specification, check your procedure and recalibrate the
sensor again.
Gas
Concentration
Pass/Fail
Air (Air is ~20.9% O2)
20.9% O2
□Yes
□No
CO
2
Verification Gas
□Yes
□No
O
2
Verification Gas
Can use Air
□Yes
□No
C2H4 Verification Gas
□Yes
□No
For information on sourcing known gases required for
calibration, please refer to the following website to inquire
about your region:
https://www.airliquide.com/group/where-we-operate