F-960 Instruction Manual rev. 6/23/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
40
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
5.
After reconnecting the sensors, insert fully charged
batteries and allow 24 hours to stabilize the
sensors.
Replacing the Oxygen (O₂) Sensor
The oxygen sensor has a life span of one (1) year, and the
replacement of the sensor is simple and quick. To purchase the
sensor from Felix Instruments contact
. To replace the sensor refer to
images on following page:
6.
Turn off the F-960 Ripen It! and remove the bottom
rubber mat.
7.
Remove the battery cap (it’s spring loaded!) and
unscrew the black bottom plate of the F-960.
8.
Unplug the red and black cable connector by gently
pulling the fastener out the end of the O₂ sensor.
9.
Unscrew the O
2
sensor (counter-clockwise).
10.
Screw in the new sensor (clockwise) until you feel a
resistance
—
not too tight!