![FE FSH Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/fe/fsh/fsh_instruction-manual_3753975056.webp)
- 48 -
INF-TN1FSH-E
T
ransmitter
Sensor 3
Sensor 2
Sensor 1
Sensor 4
Direction of flow
To line 1
To line 2
2-path system (Z method)
*4) Select sensor constant only when the sensor type is selected from FSW12, FSW21, FSW40, and FSW50.
Operation
(example)
When hybrid is selected as measurement method, Z method as sensor mount, FSW12 as sensor type,
102% for sensor constant line 1-F: (METAL), 101% for 1-R: (METAL), and 160Vpp as transmission
voltage
(When “Hybrid” is selected as measurement method, and “1 path” is selected as measurement mode)
Key operation
Description
Display
FUNC SYSTEM
Display SYSTEM.
UNIT & LANGUAGE
SKIP
"
or
#
Select “MEAS. METHOD”
MEAS. METHOD
TRANSIT TIME
ENTER
"
or
#
ENTER
Enter select/enter mode, select “HYBRID,”
and press ENTER.
MEAS. METHOD
HYBRID
#
Select “SENSOR MOUNT”
SENSOR MOUNT
V METHOD
ENTER
"
or
#
ENTER
Enter select/enter mode, select “Z METHOD,”
and press ENTER.
SENSOR MOUNT
Z METHOD
#
Select “SENSOR TYPE”
SENSOR TYPE
FSW21
ENTER
"
or
#
ENTER
Enter select/enter mode, select “FSW12,” and
press ENTER.
SENSOR TYPE
FSW12
#
Select “SENSOR CONSTANT.”
SENSOR CONSTANT.
SKIP
ENTER
"
or
#
ENTER
Enter select/enter mode, select “SETTING,”
and press ENTER.
LINE 1-F : METAL
100.00%
ENTER 1 0 2 ENTER
Enter numeric value enter mode, enter “102”
using ten keys, and press ENTER.
LINE 1-F : METAL
102.00%
"
or
#
Select “LINE 1-R”
LINE 1-R : METAL
100.00%
ENTER 1 0 1 ENTER
Enter numeric value enter mode, enter “101”
using ten keys, and press ENTER.
LINE 1-R : METAL
101.00%
ESC
Return to “SENSOR CONSTANT.”
SENSOR CONSTANT.
SKIP
#
Select “TRANS. VOLTAGE”
TRANS. VOLTAGE
80 Vpp
ENTER
"
or
#
ENTER
Enter select/enter mode, select “160,” and
press ENTER.
TRANS. VOLTAGE
160 Vpp
ESC
Display the measurement, reflecting the
setting.
(Measurement display screen)