Measurement Screen
2-5-3
0HDVXUHPHQWUDQJHDWȝV
*
>&'$PHQX&URSȝV@
:KHQWKHVDPSOLQJSHULRGLVVHWWRWKHIDVWHVWVHWWLQJRIȝVWKHPHDVXUHPHQWUDQJHZLOOEHFRPHQDUURZHU
Select the measurement range at this time.
Measurement range
CDX-L [W]15
CDX- [W] 30
CDX- [W] 85
CDX- [W]150
Near
14.0 to 14.6 mm 25.0 to 28.1 mm
22.5 to 24.0 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
65.0 t o 77.7 mm
71.5 to 74.3 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
110.0 to 134.4 mm
Center
(Initial value)
14.4 to 15.4 mm 27.8 to 31.9 mm
22.8 to 27.9 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
73.5 to 90.8 mm
70.6 to 86.9 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
124.8 to 166.3 mm
Far
15.3 to 16.0 mm 31.1 to 35.0 mm
26.7 to 28.5 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
84.8 to 105.0 mm
81.0 to 91.5 mm (with specular
UHÀHFWLRQLQVWDOODWLRQ
150.2 to 190.0 mm
2-5-4 Upper Limit/Lower Limit
Select “Auto” for the sampling period to set the maximum and minimum sampling periods to adjust.
The actual sampling period will be automatically adjusted within this set range.
ź8SSHU/LPLW
>ȝVHF@
Default value: 500
ź/RZHU/LPLW
>ȝVHF@
Default value: 25
2-5-5 Measuring
object
Selects the target to measure.
Normally select “Standard.”
Refer to the table shown below and select “Thin Glass” when measuring the surface displacement or thickness
of a thin glass plate or a similar object.
* The standard thickness varies depending on the material of the target.
* When you use “Thin Glass”, the linearity becomes 5 to 10 times worse.
CDX-L [W]15
CDX- [W] 30
CDX- [W] 85
CDX- [W]150
Standard thickness
0.2 mm
0.7 mm
3.0 mm
Cannot be measured
ź0HDVXULQJREMHFW
Standard/Thin Glass
Default value: Standard