Chapter 2: CAM2® Measure Devices Menu
coordinate system. After the probe's coordinates are established, these
are translated into FaroArm coordinates and you are ready to start taking
measurements.
Measurement accuracy relies on probe calibration under optimal
conditions. If the probe calibration passes, measurements will be
accurate. If the probe calibration fails, measurements will not be as
accurate. Proper mounting and technique are critical calibration factors.
To optimize calibration and minimize
stress-induced errors during this critical
procedure, place the FaroArm in a single
position in which the elbow joint
remains relatively stationary without any
obstructions or restrictions in movement
while the calibration is performed. Do
not let the elbow joint fall during
calibration. Only exercise the last joint
of the FaroArm for the software to
accurately calibrate the probe tip.
Minimizing elbow joint movement and
focusing on the probe and last arm joint
during calibration maximizes
measurement accuracy.
Calibrate Probe
When changing the style or dimension of the probe at the end of the
FaroArm, you must calibrate it for the FaroArm to measure and function
accurately. Probe calibration is an optimization procedure that requires
you to digitize points. The two methods of calibration are:
• Single Hole
• Sphere
Click either button to calibrate the current probe. The Single Hole
Calibration method differs depending on whether the G
UIDANCE
check
box is selected or cleared.
•
See “Single Hole Method - Guidance” on page 80.
•
See “Single Hole Method” on page 83.
•
See “Sphere Method” on page 85.
08M46E00_FaroArmUSB.book Page 79 Tuesday, February 17, 2009 1:42 PM
Содержание FaroArm Fusion
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Tuesday February 17 2009 1 42 PM ...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Tuesday February 17 2009 1 42 PM ...
Страница 8: ...08M46E00_FaroArmUSB book Page iv Tuesday February 17 2009 1 42 PM ...
Страница 100: ...08M46E00_FaroArmUSB book Page 92 Tuesday February 17 2009 1 42 PM ...
Страница 113: ...B 9 08M46E00_FaroArmUSB book Page 9 Tuesday February 17 2009 1 42 PM ...
Страница 114: ...08M46E00_FaroArmUSB book Page 10 Tuesday February 17 2009 1 42 PM ...
Страница 120: ...08M46E00_FaroArmUSB book Page 6 Tuesday February 17 2009 1 42 PM ...
Страница 129: ...08M46E00_FaroArmUSB book Page 1 Tuesday February 17 2009 1 42 PM ...