FaroArm records points without pressing either button.
• Re-seat the set screw in the FaroArm handle. If the set screw is not
seated correctly it causes intermittent contact to the pin underneath,
and random points are taken.
1 Remove any standard FaroArm probe.
2 Remove the set screw using the supplied hex wrench.
3 Replace the screw.
N
OTE
: Be careful not to bottom out the screw. This may cause the
pin underneath to be constantly depressed possibly damaging the
FaroArm.
4 Replace the FaroArm probe.
Power Issues
If the FaroArm is receiving power from the power source, or outlet, but
not powering up the system, the source of this issue may lie in one or all
of the following areas:
• Ensure you are using the correct cables.
• Check the cables and connectors for bent pins or damage and ensure
that the connectors are secure. Replace all cables that show signs of
damage.
• Check the power source (outlet) for proper voltage, current and
grounding. For more information, see “FaroArm Power Supply” on
page 56.
Figure 1-43 Handle Screw
08M46E00_FaroArmUSB.book Page 60 Tuesday, February 17, 2009 1:42 PM
Содержание FaroArm Fusion
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Tuesday February 17 2009 1 42 PM ...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Tuesday February 17 2009 1 42 PM ...
Страница 8: ...08M46E00_FaroArmUSB book Page iv Tuesday February 17 2009 1 42 PM ...
Страница 100: ...08M46E00_FaroArmUSB book Page 92 Tuesday February 17 2009 1 42 PM ...
Страница 113: ...B 9 08M46E00_FaroArmUSB book Page 9 Tuesday February 17 2009 1 42 PM ...
Страница 114: ...08M46E00_FaroArmUSB book Page 10 Tuesday February 17 2009 1 42 PM ...
Страница 120: ...08M46E00_FaroArmUSB book Page 6 Tuesday February 17 2009 1 42 PM ...
Страница 129: ...08M46E00_FaroArmUSB book Page 1 Tuesday February 17 2009 1 42 PM ...