Chapter 1: Introduction to the FaroArm
®
When sweeping from side to side and front to back, make sure you
sweep through at least 90 degrees.
Typically, you should take about 200 scans during the calibration
process.
In the probe calibration log, look at the X, Y, Z, A, B, C values. Each
value should repeat to within:
X: 100 microns A: 0.1 degrees
Y: 50 microns B: 0.05 degrees
Z: 50 microns C: 0.1 degrees
N
OTE
: Values are for V2 Laser Line Probe
If additional support is required, please contact customer support. When
contacting customer service, please provide the following information:
1 The serial number of the device(s).
2 A copy of the probe calibration log.
3 A copy of the event log.
4 A copy of the Single Point Articulating Test (SPAT).
5 Pictures of the setup.